3.3.5 Calibration of Vacuum System Mass Spectrometers
365
research, development, or process reliability. Several techniques that can be used
to calibrate vacuum instrumentation are presented next. Detailed descriptions of
these methods are provided in Herman [31] and Ellefson and Cain [32].
3.3.5.1
Static Volume Expansion
The static volume expansion simply involves establishing a pressure that is mea-
surable by a primary standard within a very small volume and expanding that vol-
ume of
gas
into a much larger volume to obtain a substantial reduction in pressure
(Boyle's law). Figure 11(a) is a schematic of the technique. The small volume, V^,
typically a few cm\ and the large volume V2, typically > 10^ cm\ are initially
cleaned and evacuated to the UHV range to minimize outgassing contributions to
the signal. The small volume is then pressurized topi = 10"^ or 10~^ torr with
the separation valve closed. The small-volume pressure pi is measured with a
primary standard or a working standard such as a fused silica Bourdon tube , a
capacitance manometer, or a spinning rotor gauge and then the separation valve
(which has a large-diameter bore) is fully opened. The equilibrium pressure then
becomes
PJ = (^4)
Vacuum chamber wall pumping effects such as chemisorption and/or incor-
poration can be serious problems and several runs must be made to saturate the
walls and obtain repeatability. Furthermore, if hydrogen is the calibration gas, the
walls must be significally oxidized to inhibit the seemingly endless hydrogen in-
corporation. This surface condition can be achieved with a bakeout of ~400°C
in an atmosphere of oxygen for several hours. Thicker oxides, of course, can be
obtained using longer times and higher temperatures. Another concern is gauge
pumping. The ordinary ion source in a mass spec operating at an emission current
of ~1 mA may have a pumping speed for N2 approaching 0.1 L/s, which can
have a substantial effect on the pressure diminution. Extrapolation techniques can
be used to estimate the correct expanded pressure, but this is not a totally accept-
able procedure. If wall effects and the gauge pumping are not serious problems,
calibrations as low as 1 X
10 ~^
torr , depending on the gas, can be achieved. If
the two volumes are initially degassed to a base pressure of
1
X 10"^^ torr, wall
outgassing in the static system in V2 may increase the background pressure
two-three decades over the time of the individual measurements. Thus, there is a
1%
or greater uncertainty in the calibration gas. In practice, the low-pressure cali-
bration limit is probably 1 X 10"^ torr. This technique is not generally used be-
cause of the aforementioned effects and because of its one-decade-or-less range
overlap with ionization-type instruments.