748 Chapter 5.6: Common Analytical Methods for Surface and Thin Film
Unlike the electron spectroscopies, SIMS is not considered an "inherently"
quantitative method. However, using appropriate standards in the dynamic mode,
SIMS can be very quantitative. In some of
the
early quantitation reported, samples
were implanted with the atom to be quantified. The implantation method yields a
quantitative amount of implanted atoms/cm^ (measured charge) and knowing the
implantation energy (measured accelerating potential), the ion penetration depth
can be obtained from tables [18]. Implants are Gaussian distributed to first order.
Consequently, the peak of the implant profile as determined from the standard
provides the concentration in a known matrix. This method is described in detail
in Reference [19]. SIMS can also provide isotopic information. For example, hy-
drogen and deuterium are readily separated by most spectrometers.
5.6.3.2 Instrumentation: Dynamic SIMS
Dynamic SIMS is used primarily for depth-profiling thin-film structures. In addi-
tion to the ion source, this instrument requires an ion extraction lens close to the
sample surface, a low-energy electrostatic energy analyzer, and a quadrupole mass
spectrometer or other mass spectrometer. Larger ion current densities must be
used to provide continuous sputtering, and in some cases very-high-rate material
removal (A per second) can be obtained using Cs, Ar, O, and Ga beams. The vari-
ous sputtering ions change the sensitivity and ion yield. The ion beam is rastered
over a region of interest on the sample surface, and the secondary ion signal is
raster gated in a smaller region to eliminate or minimize crater edge effects, as
shown earlier in the case of elemental depth profiling with AES. In this mode of
operation, very good depth resolution can be obtained over large film thicknesses
Ion signals are analyzed in a multiplex mode. Instruments are computer con-
trolled, and the computer sets the amount of time spent per mass channel. The re-
sulting profile is plotted as the number of counts in a channel as a function of
sputter time. As noted earlier, the counts scale can be changed into a quantitative
scale. The depth scale can be measured directly, using a surface stylus instrument
after the sample has been sputtered. Conversion to concentration versus depth is
therefore quantitative for simple materials systems.
Elemental depth profiling is generally performed over many thousand ang-
stroms of material. Depth resolution may be very material dependent. For ex-
ample, if the sample consists of palladium on silicon, it is likely that atomic mix-
ing will occur as a result of the cascade energy transfer during profiling. In this
case the depth resolution would be poor. Resolutions of less than 200A over a
depth of several tens of micrometers have been routinely achieved in other mate-
rial systems.
Dynamic systems have been used to study ion implantation profiles with sensi-
tivities over five orders of magnitude in crystalline materials such as silicon or