Solid-State Dosimeters 399
Figure 8.64a shows the sensitivity curves for four detec-
tors (0.5 mm, 0.3 mm, 0.15 mm a-Se on 2-mm Al, and
0.3 mm a-Se on 1-mm Cu) irradiated with the
60
Co spec-
trum. The uncertainties of all the values in Figure 8.64a are
less than 0.9%; thus, the error bars are too small to be
shown.
There is no significant difference between the sensi-
tivity curves of the detectors with Al (0.54 g/cm
2
) or Cu
(0.89 g/cm
2
) metal plates and which have the same a-Se
thicknesses. The relative
values were calculated by the
Monte Carlo method. The D
m
values produced the best fit
of Equation (8.26) to the data in Figure 8.64a. The corre-
sponding correlation coefficients R are at least 0.999 for
all fittings. Once D
m
is obtained, the parameter a is then
calculated from Equation (8.25).
Detector sensitivity was also measured for the 6-MV
spectrum. In Figure 8.64b, the discharge curves for seven
detector types are shown. The uncertainties of all the
values in the figure are less than 1.2%.
The sensitivity curve of the metal/a-Se detectors was
measured with constant electric field, and results are shown
in Figure 8.64c. The largest values for
V
grid
and |V
bias
| used
in our experiments were 2100 V and 2000 V, respectively.
Since the resultant electric field (E) across the largest a-Se
thickness (0.50 mm) was 8.2 V/
m, this constant value
of E was used for all studies. For the a-Se thicknesses less
than 0.50 mm, smaller values of V
grid
and V
bias
were
required to achieve the same E. Thus, potentials V
0
of 1230
V, 2460 V, and 4100V were placed on detectors with a-Se
layer thicknesses of 0.15 mm, 0.30 mm, and 0.50 mm,
respectively. [32]
REFERENCES
1. Vatnitsky, S. and Jarvinen, H., Phys. Med. Biol., 38,
173, 1993.
2.
Hoban, P. W. et al., Phys. Med. Biol., 39, 1219, 1994.
3.
Rikner, G., Acta Radiologica Oncol., 24, 71, 1985.
4.
Seuntjens, J. et al., in AAPM Proc. No. 11, Kilovolt
X-Ray Beam Dosimetry for Radiotherapy and Radiobiol-
ogy,
1997, 227.
5.
Heydarian, M. et al., Phys. Med. Biol., 38, 1035, 1993.
6.
PTW-Freiburg Ionization Chamber Catalog, 1999.
7.
Vatnitsky, S. et al., Med. Phys., 22, 469, 1995.
8.
Rustgi, S. N., Med. Phys., 22, 567, 1995.
9.
Laub, W. U. et al., Med. Phys., 24, 535, 1997.
10.
Rustgi, S. N., Phys. Med. Biol., 43, 2085, 1998.
11.
Mobit, P. N. and Sandison, G. A., Med. Phys., 26, 839,
1999.
12.
Gladstone, D. J. et al., Med. Phys., 21, 1721, 1994.
13.
Soubra, M. et al., Med. Phys., 21, 567, 1994.
14.
Butson, M. J. et al., Med. Phys., 23, 655, 1996.
15.
Rosenfeld, A. B. et al., IEEE Trans. Nucl. Sci., 42, 1870,
1995.
16.
Rosenfeld, A. B. et al., IEEE Trans. Nucl. Sci., 43, 2693,
1996.
17.
Korn, T. et al., Phys. Med. Biol., 43, 3235, 1998.
18.
Rosenfeld, A. B. et al., IEEE Trans. Nucl. Sci., 46, 1774,
1999.
FIGURE 8.64 Radiation discharge curves for different metal
plate/
a-Se detectors measured with the (a)
60
Co spectrum,
(b) 6-MV spectrum using
V
grid
2100 V and V
bias
0, and (c)
radiation discharge curves for detectors having the same electric
field
E 8.2 V/
m across the a-Se layer. The same E was
obtained by applying
V
0
1230 V, 2460V, and 4100V for the
detectors with
a-Se layer thicknesses of 0.15 mm, 0.30 mm, and
0.50 mm, respectively.(From Reference [32]. With permission.)
Ch-08.fm Page 399 Friday, November 10, 2000 12:03 PM