382 Radiation Dosimetry: Instrumentation and Methods
oxide charge to the interface. The sensitivity to tempera-
ture can be minimized by operating the p-FET with a
current at the temperature-independent point. The main
limitation in using silicon semiconductor detectors is
effects caused by radiation damage. A nonlinear response
with dose rate has proved to be serious for n-type detectors
when pre-irradiated to 10 kGy or more in 20-MeV elec-
trons. A sensitivity drop after pre-irradiation is another
effect which is more pronounced for n-type detectors com-
pared to p-type detectors of the same resistivity. A third
effect of radiation damage is the sensitivity increase with
increasing temperature. [20]
A Pantak HP-320 quasi-monoenergetic x-ray unit was
used by Edwards et al. [21] to determine the response
of two Thomson and Nielson TN-502RD MOSFETs, a
Scanditronix EDP-10 semiconductor (build-up cap 10 mm:
tissue equivalence), a BDD-5 semiconductor (build-up
cap 4.5 mm: tissue equivalence), and a LiF:Mg:Ti TLD
over the energy range 12–208 keV. The sensitivity of
each detector was normalized to the value produced by
exposure to 6-MV x-rays. The maximum relative sensi-
tivities of the two MOSFET detectors were 4.19
0.25
and 4.44 0.26, respectively, occurring at an incident
x-ray energy of 33 keV. The maximum relative sensitiv-
ity of the Scanditronix EDP-IO of 2.24 0.13 occurred
at 65 keV, and for the EDD-5, it was 7.72 0.45 at 48 keV.
The TLD produced a maximum relative sensitivity of 1.31
0.09 at 33 keV. Compared with available data based on
heteroenergetic x-ray sources, these measurements have
identified a more representative response for each detector
to low-energy x-rays.
Scanditronix
p-type semiconductor detectors were
investigated (Scanditronix, Uppsala, Sweden). These
included the EDD-5 and the EDP-IO, both connected to
a DPD-510 direct patient dosimeter. The 0.9-mm sensitive
volume of the EDD-5 detector is encapsulated by Perspex,
which provides an effective tissue-equivalent depth of
4.5 mm. This detector is designed primarily for critical
organ measurements outside the main beam. The 0.9-mm
3
sensitive volume of the EDP-IO detector is surrounded by
a 0.75-mm-thick stainless steel cap (providing a tissue-
equivalent depth of 10 mm) and is adhered to a Perspex
plate (1-mm thickness). This detector is designed for mea-
surements inside the main beam generated by 4–8 MV
linear accelerators.
A correction factor U
d
for each detector d was calcu-
lated to allow for the small difference between the SSD
to the center of its sensitive volume (J
d
) and the SSD to
a rigid surface:
(8.10)
The sensitivity
S
X,d
of a detector at each x-ray energy
relative to that produced by 6-MV x-rays was calculated
from:
(8.11)
where
R
x,d
is the detector response, b
d
is the response to
background radiation, k
x
is the kerma rate recorded by the
NE2550 Protection Level Secondary Standard with a beam
current of 1 mA and SSD of 1 m, and t
x
is the exposure time.
The corrected response per unit air-kerma rate was normal-
ized to the detector’s response per unit air kerma rate from a
6-MV linear accelerator by multiplying by two previously
determined calibration factors. The first factor L
d
converted
each detector’s response to the absolute dose rate, measured
at the maximum depth dose in a water phantom irradiated at
an SSD of 1 m with a 10-cm 10-cm field from a 6-MV
linear accelerator. The second factor
M
d
converted this to the
corresponding air-kerma rate and included any effects due to
attenuation within the detector and from backscatter.
The variation in relative sensitivity of each detector with
x-ray energy is illustrated in Figure 8.35. Relative to its
response to a 6-MV linear accelerator beam, the sensitivity
of each detector increased as the x-ray energy decreased
until it reached a maximum value, and then it decreased
with further decrease in x-ray energy.
On-chip p-FETs were developed by Buehler et al. [22]
to monitor the radiation dose of n-well CMOS ICs by
monitoring threshold voltage shifts due to radiation-
induced oxide and interface charge. The design employs
closed-geometry FETs and a zero-biased n-well to elimi-
nate leakage currents. The FETs are operated using a
constant current chosen to greatly reduce the FETs’ tem-
perature sensitivity. The dose sensitivity of these p-FETs
is about 2.6 mV/krad(Si) and the off-chip instrumenta-
tion resolves about 400 rad(Si)/bit. When operated with a
current at the temperature-independent point, it was found
that the pre-irradiation output voltage is about 1.5 V,
which depends only on design-independent silicon mate-
rial parameters. The temperature sensitivity is less than
63
V/°C over a 70°C temperature range centered about
the temperature insensitive point.
The schematic cross section of the device shown in
Figure 8.36 indicates that the n-well and source are sep-
arated so that they can operate at slightly different biases
required by the operational amplifier, U
l
. This allows all
the forced current to flow through the p-FET channel.
Drain-to-well leakage is shunted to ground. Grounding
the n-well is a departure from normal CMOS circuit oper-
ation, where the n-well is normally connected to VDD.
In operation, all terminals of the p-FET are operated
near ground except the drain, which operates near V0
1.5 V. During irradiation, the device is biased in the off state.
The two RL resistors are used to bleed off any charge
U
d
0.4
J
d
-------
2
S
x,d
L
d
M
d
R
x,d
b
d
()
k
x
t
x
A
x
F
x
U
x
-----------------------------------------
Ch-08.fm Page 382 Friday, November 10, 2000 12:03 PM