Film Dosimetry 319
method to calculate the dose response curves (net optical
density at a given wavelength or spectrum vs. absorbed dose)
for different densitometer light sources using measured GC
film absorption spectra. Comparison with measurements on
the latest version of GC film (Model MD-55-2) using four
types of densitometers [He-Ne laser, broadband (white
light) densitometer, and two LED (red-light) filtered den-
sitometers] confirm the accuracy of this predictive model.
They found that an LED (red-light) source with a narrow
band-pass filter centered at 671 nm near the major absorp-
tion peak achieves nearly the maximum possible sensitiv-
ity (almost four times more sensitive than He-Ne laser,
632.8 nm) and may be suitable for in vivo dosimetry.
For LED light source coupled to a band-pass filter,
the resulting intensity as a function of wavelength after
having passed through the filter is
(5.14)
where I
0
(
) is the intensity of the light source (LED) at
wavelength
incident on the filter; I
1
(
) is the intensity
after having passed through the filter; and A
filter
(
) is the
absorbance of the filter at wavelength
. For the case of
the broadband (white light) light source, I
0
(
) was deter-
mined from the temperature of the tungsten filament, an
assumed blackbody radiation spectrum, and, for mathe-
matical convenience, A
filter
(
) 0 (no filter). By coupling
the light source
I
1
(
) with the absorption spectra of
GafChromic film, the intensity reaching the detector can
be written as
(5.15)
where
A
GAF
(
, D) is the absorbance of GafChromic film
at wavelength
, irradiated to a dose D. Optical density
as a function of dose [OD(D)] for GC film was calculated
by combining the equations above to yield
(5.16)
By subtracting the optical density of an un-irradiated
sample of GafChromic film, the net optical density (NOD)
as a function of dose was obtained (dose response curve)
for each densitometer light source. Note that the model
assumes a uniform response by the detector for the range
of wavelengths measured. The transmission spectra for
the two band-pass filters used in our measurements with
the filtered LED densitometers are shown in Figure 5.28.
Absorption spectra of GC film exposed to total doses 0,
10, 20, 30, 50, and 100 Gy and measured at 31.9 days
post-irradiation are shown in Figure 5.29. The minor and
major absorption peaks can clearly be seen. At 30 Gy
exposure, these peaks are centered at 614 nm and 674 nm,
respectively, with peaks shifting to lower wavelengths
with increasing dose.
Figure 5.30a plots the predicted dose response curves
at fixed wavelengths 400 nm, 510 nm, 632.8 nm, 650 nm,
and 671 nm as calculated from the measured absorption
spectra of GC film (MD-55-2). It is observed that the dose
response curve measured at 671 nm (near the major peak)
is the most sensitive with a DNODI 14 Gy. The least
sensitive dose response curve is at 400 nm (not near either
of the absorption peaks), where a net optical density of
1 is not achieved, even for a dose as high as 100 Gy. Of
the fixed wavelengths, the dose response curve at 650 nm
(on the left shoulder of the major peak) exhibits the great-
est linearity, with a DNODI 34 Gy. Also shown is the
dose response curve for a fixed wavelength of 632.8 nm,
which can be compared with the measured data from the
He-Ne laser densitometer. The predicted and measured
results show excellent agreement (DNODI 56 Gy for
both). [12]
The dose response of high-sensitivity GafChromic
film to photons from
125
I seeds for doses up to 200 Gy
FIGURE 5.27 A schematic view of the structure of GafChromic
MD-55, indicating the seven layers and their thicknesses. The
uncertainty in the measurement of the thicknesses was estimated
to be about
0.001 mm for the hard Mylar layers and 0.002
mm for the soft, gel, and glue layers. (From Reference [11]. With
permission.)
I
1
() I
0
()10
A
filter
()
I
2
, D()I
1
()10
A
GAF
,D()
OD D() log
I
0
()10
A
filter
()
10
A
GAF
,D()
0
dI
0
()10
A
filter
()
d
0
Ch-05.fm Page 319 Friday, November 10, 2000 12:01 PM