130 Radiation Dosimetry: Instrumentation and Methods
The radiation exposure from an x-ray tube can be
measured by an ion chamber as
(3.81)
(3.82)
where
X is exposure measured in C/kg, M is meter reading
in C, N
X
is the calibration factor (R/C) of an ion chamber
for standard beam energy from a national or accredited
dosimetry calibration laboratory (ADCL), K is a compos-
ite correction factor, k
e
is the energy correction factor, k
tp
is the correction for the air density due to the temperature
and pressure, k
pol
is correction for the polarity, and P
ion
is
the ion recombination correction factor. An ADCL pro-
vides values of N
X
for each user’s beam quality condition
for an ion chamber and electrometer combination. How-
ever, the dose rate might be a variable and the ADCL may
not be able to match the user’s exposure condition. Even
though for routine use, a single factor might be useful,
scientifically, separation of each component of the cali-
bration factor individually as shown in Equation (3.82) is
desired. In such situations, an ADCL would prefer to
provide individual factors for each beam condition. Ide-
ally, ionization chambers should be independent of an
electrometer such that chamber factors can be provided
by an ADCL very similar to the factors for chambers used
in radiation therapy applications. If a chamber is cali-
brated by an ADCL with a polarity that is used by a
user,
k
pol
is 1.00 and further correction is not needed.
However, if calibration and measurement polarities are
different, k
pol
is needed for accurate exposure measure-
ment.
For an x-ray machine, the exposure is dependent upon
the applied voltage, V; beam current, mA (milliampere);
time, s (seconds); and distance, d. A rule of thumb expres-
sion for exposure can be given as: [45]
(3.83)
where k and n are constants for a machine. The typical
values of k and n are 1.5 and 2.5, respectively, in the
kilovoltage range. Hence, the polarity and ion recombina-
tion effects in general may depend on these parameters.
For the continuous radiation beam, the ion-recombi-
nation correction P
ion
can be calculated by the half-voltage
method as suggested by Attix. If Q
1
and Q
2
are the charges
collected at the chamber potential of V and V/2, respec-
tively, P
ion
is calculated as
(3.84)
The polarity effect depends strongly on the medium
between the plates of chamber and is the highest for air
( 30%) and the lowest ( 0.3%) for dielectric medium.
If the charges collected at negative and positive potentials
are denoted by Q
and Q
, respectively, then k
pol
is defined
as the absolute value of the ratio.
(3.85)
The actual measured charge Q for the measurement of
exposure is the absolute value of mean of the two charges:
(3.86)
For the measurements, if only a single polarity is used,
the error in
Q will be
(3.87)
(3.88)
The P
ion
and k
pol
were measured with various exposure
factors kVp, mA, s, and d for all chambers in the diag-
nostic energy range by changing one parameter at a time
while keeping the others constant.
Figure 3.45 shows the variation of the P
ion
vs. x-ray
tube voltage in mammography range. Data are shown for
various ion chambers at a focus-to-chamber distance
(FCD) of 25 cm and 50 cm in Figure 3.45a and b, respec-
tively. The value of P
ion
is relatively independent of the
tube voltage for small-volume (150 cm
3
) ion chambers
within 1%; however, for the large-volume ion chambers,
the magnitude of P
ion
is much higher and rises steadily
with the tube potential. In general, P
ion
is greater for the
higher exposures. The slope of curve for the 180-cm
3
chamber was anomalously higher. Independent repeated
measurements with this chamber showed the same results
within 1.5%.
The effect of beam current for a given exposure (fixed
kVp and mAs) is shown in Figures 3.46a and b for the
P
ion
and k
pol
, respectively, for the 125-kVp and 0.1-s expo-
sure typically used in diagnostic radiology. The value of
P
ion
rises linearly with beam current for most chambers,
but the effect is more pronounced for the large-volume
ion chambers. For small-volume chambers, the P
ion
is con-
stant within 1.0%. It suggests that P
ion
is significant only
at high-exposure rate measurements with large chambers.
The polarity effect, on the other hand, is constant over a
wide range of beam current. The magnitude of the k
pol
is
Exposure
Kk
e
k
tp
k
pol
P
ion
R()exp
kV
n
mAs
d
2
--------------------
P
ion
4
3
---
Q
1
Q
2
------
1
k
pol
Q
Q
-------
Q
Q
Q
2
-----------------------
%error
k
pol
1
k
pol
1
------------------
100 for Q
%error
1 k
pol
1 k
pol
------------------
100 for Q
Ch-03.fm(part 1) Page 130 Friday, November 10, 2000 11:58 AM