Ionization Chamber Dosimetry 119
negative chamber bias. These readings were corrected for
the polarity effect.
Build-up curves in polystyrene for 5 5 and 20
20 cm
2
6-MV x-ray fields angled at 60° and 75° with no
blocking tray in the beam are shown in Figure 3.35. The
%DD values in the graphs are obtained by dividing the ioni-
zation charge collected at the slant depth along the central
ray of the beam by the ionization charge collected with the
same detector at the reference depth of a normally incident
beam. The relative response as a function of depth along
the central axis of the beam for the Memorial chamber and
TLD powder is also shown in Figure 3.35.
The magnitude of the over-response at the phantom
surface for several fixed-separation plane-parallel cham-
bers as a function of beam angulation for a 5 5-cm
2
field is shown in Figure 3.36. Relative response for these
figures is defined as the %DD at the surface for the test
chamber for a beam incident at
° divided by the %DD
for the extrapolation chamber, 0-mm plate separation, at
the surface for a beam incident obliquely at the same
angle. The response of the chambers at 0°, 30°, 45°, 60°,
and 75° relative to the response of the extrapolation
chamber is shown in Figure 3.36a for 6-MV x-rays. Based
on this data, the relative response of the chambers at 0°
ranged from 1.6 to as high as 2.3 times greater than the
response of the extrapolation chamber. The amount of
over-response increases gradually with increasing angle,
reaching a maximum at about 45°. At beam angles
greater than 45°, the amount of chamber over-response
begins to decline. At a beam angulation of 75°—the high-
est angle of oblique incidence investigated—the chamber
over-response ranges from 1.4 to 1.8. In comparison to
other chambers, the Attix chamber shows a relatively
small amount of over-response at 0°—a factor of about
1.12—with little variation up to a beam angulation of
75°. Angular chamber response at the surface for 24-MV
x-rays is shown in Figure 3.36b. Three of the chambers
show approximately a 1.25 over-response at 0°, while
the Markus chamber over-responds by a factor of approx-
imately 1.4. This over-response increases dramatically
with beam angulation, reaching a maximum between 1.7
and 1.95 at 60°. From 60° to 75°, chamber response
declines for the Markus and the Memorial (circular)
chamber but remains constant for the Capintec and the
Memorial (rectangular) chamber. By comparison, the
Attix chamber shows an over-response of about 1.05 at
0°, with a maximum over-response of 1.2 at 75°.
Again, the amount of over-response of the Attix chamber
is very similar to that exhibited by the extrapolation
chamber with a 2.5-mm plate separation. [35]
From the data shown, electron fluence across a plane-
parallel ionization chamber is disrupted less when the
chamber separation is small, and when the active volume
is isolated from the in-scattering effects of the side walls
of the chamber. This is not only true when high-energy
photons are normally incident upon the chamber, but also
when these beams are obliquely incident upon the chamber.
IV. CHARGE COLLECTION, ION
RECOMBINATION, AND SATURATION
The output current, i, from an ionization chamber is reduc-
ed from the saturation current, i
s
, by initial recombination,
back-diffusion to electrodes, and volume recombination.
TABLE 3.12
Effect of Density of Chamber Walls (
60
Co; 10
10-cm
2
field at 80 cm; 5-mm-diameter collector)
Electrode
Separation
(mm)
Over-response
Wall
Density
Wall
Diameter
(mm)
For
Indicated
density
For
Density
1.17
0.22 12 3.5 5.8% 17.0%
0.22 22 4.1 3.7% 10.3%
0.22 12 2.0 2.2% 9.7%
0.22 22 2.0 1.3% 5.0%
2.15 12 3.1 26.5% 15.1%
2.15 22 3.1 14.0% 7.8%
Source: From Reference [34]. With permission.
FIGURE 3.34 A suggested design of a fixed-separation parallel-plate ion chamber giving acceptably small side-wall error. (From
Reference [34]. With permission.)
Ch-03.fm(part 1) Page 119 Friday, November 10, 2000 11:58 AM