Ionization Chamber Dosimetry 97
Both IAEA TRS-381 and the new UK-protocol
(IPEMB) [12] define h
m
as the ratio of electron fluences
at equivalent depths in the water and plastic phantoms. If
a criterion is applied where perturbation factors in water
and plastic differ, the experimental determination of the
factor h
m
should strictly be modified according to
(3.21)
where the ratio of perturbation factors for the chamber
used in the measurements appears explicitly for the two
media. is the ratio of the electron fluences as noted
by AAPM Task Group 21. [13]
AAPM Task Group 39 [14] has presented a different
approach. Instead of calculating the dose to water directly,
these protocols determine first the dose to the plastic phan-
tom using the equation
(3.22)
where M is the electrometer reading in the solid plastic
phantom, corrected for pressure and temperature. is
the cavity gas calibration factor, is the mean
restricted mass stopping power of plastic to gas, P
ion
the
correction for ion recombination, P
repl
the perturbation
replacement factor, and P
wall,plastic
the perturbation wall fac-
tor in the plastic medium.
The dose to water is then obtained in a second step
by scaling the electron fluence in water and the solid
phantom by the relation [11]
(3.23)
where is the fluence factor, i.e., the ratio of electron
fluence in water to that in the solid phantom at equivalent
depths, and ( )
coll
is now the mean unrestricted mass
stopping power ratio.
Figure 3.9 shows the measured P
wall
factors as a func-
tion of energy for the simulated Attix, Roos, and NACP
chambers placed in phantoms of aluminum (Figure 3.9a)
and polyethylene (Figure 3.9b). These materials are not
typical for clinical dosimetry but have atomic numbers
that differ significantly from the chamber wall materials
and thus illustrate wall effects more clearly. [11]
Figure 3.10 shows the P
wall
-factors for graphite,
PMMA, clear polystyrene and Plastic Water™. These
phantom materials are closer to the wall materials with
respect to atomic number and scattering effects and, there-
fore, the P
wall
-factors are also closer to unity. The NACP
chamber is recommended to be used with a PMMA phan-
tom. At low electron energies where plastic phantoms are
normally used, it can be seen in Figure 3.10 that the
P
wall
-
factor is 0.99. The Roos chamber has a P
wall
-factor equal
to unity in a PMMA phantom. Figure 3.10 also shows that
the P
wall
-factor for the Attix chamber in PMMA is close
to 1.00, as the backscatter in PMMA and Solid Water™
are similar. In a clear polystyrene phantom, the P
wall
-fac-
tors are below unity for all chambers, especially for the
NACP chamber, where a P
wall
-factor of 0.97 was obtained
at low energies. Also, the Attix and Roos chambers have
P
wall
-factors below 0.99 at low energies.
B. k
Q
FACTORS
k
Q
, the beam quality correction factor, corrects the
absorbed dose-to-water calibration factor N
D,w
in a refer-
ence beam of quality Q
0
to that in a user’s beam of quality
Q
1
. It was discussed by Vatnitsky et al. [15]
If an ionization chamber with a
60
Co absorbed dose-
to-water calibration (reference beam, quality Q
0
) is used
for measurements in a beam with quality Q
1
, the absorbed
dose-to-water D
w
(Q
1
) is given by (rewriting Equation 3.3)
(3.24)
where
M
Q
is the charge collected in the ionization chamber
in the beam Q
1
corrected to reference temperature and
pressure conditions and for ionic recombination; N
D,w,
is
the absorbed dose-to-water calibration factor in the
60
Co
beam; and k
Q
corrects the reference absorbed dose-to-
water calibration factor N
D,w,
to the beam with quality Q
1
.
The subscript
is placed on k
Q
to specify that the refer-
ence beam is the
60
Co beam. The term ‘‘quality” is used
to specify the radiological properties of the beam.
For photons, the quality is specified in terms of the
and, for electrons, in terms of the depth of 50%
dose in water. The beam quality specifier for a proton
beam is the effective energy of the protons at the calibra-
tion point.
Water calorimetry was used as the absolute dose stan-
dard to measure k
Q
, employing the water calorimeter devel-
oped by R. Schulz. [16]
The dose delivered to the water for irradiation within
a reference beam of quality Q
0
(
60
Co beam) was deter-
mined from
(3.25)
where
c is the specific heat of water at the calorimeter
operation temperature,
is the calibration factor obtained
from the thermistor calibrations, V
0
is the bridge deflec-
tion, and D
T
is the thermal defect.
Vatnitsky et al. determined k
Q
by treating the calo-
rimeter response to the irradiation as a deflection of the
Wheatstone bridge. When the water calorimeter was
replaced with a water phantom, the absorbed dose to water
measured by an ionization chamber in the same beam Q
0
was given by
(3.26)
h
m
exp
med
water
p
Qmed,
p
Qw,
med
water
D
med
MN
gas
pp
L
()
gas
med
P
ion
P
repl
P
wall plastic,
N
gas
pp
L
()
D
med
d
water
()D
med
d
med
()S
()
coll
[]
med
water
med
water
med
water
S
D
w
Q
1
() M
Q
1
N
Dw
,,
k
Q
TPR
10
20
D
water
Q
0
cal,()c
V
0
1 D
T
()
D
water
Q
0
ion,()N
Dw
,,
M
0
Ch-03.fm(part 1) Page 97 Friday, November 10, 2000 11:58 AM