
© 1999 by CRC Press LLC
measurements to minimize effects of friction and other lateral forces and to measure topography of soft
surfaces. The tapping mode is also referred to as dynamic force microscopy.
The multimode AFM, used with a grounded conducting tip, can measure electric field gradients by
oscillating the tip near its resonant frequency. When the lever encounters a force gradient from the electric
field, the effective spring constant of the cantilever is altered, changing its resonant frequency. Depending
on which side of the resonance curve is chosen, the oscillation amplitude of the cantilever increases or
decreases due to the shift in the resonant frequency. By recording the amplitude of the cantilever, an
image revealing the strength of the electric field gradient is obtained.
In its simplest form, MFM used with a magnetically coated tip detects static cantilever deflection that
occurs when a magnetic field exerts a force on the tip, and the MFM images of magnetic materials can
be produced. Multimode AFM enhances MFM sensitivity by oscillating the cantilever near its resonant
frequency. When the tip encounters a magnetic force gradient, the effective spring constant, and hence
the resonant frequency, is shifted. By driving the cantilever above or below the resonant frequency, the
oscillation amplitude varies as the resonance shifts. An image of magnetic field gradients is obtained by
recording the oscillation amplitude as the tip is scanned over the sample.
Topographic information is separated from the electric field gradients and magnetic field images by
using a so-called lift mode. Measurements in lift mode are taken in two passes over each scan line. On
the first pass, topographical information is recorded in the standard tapping mode where the oscillating
cantilever lightly taps the surface. On the second pass, the tip is lifted to a user-selected separation
(typically 20 to 200 nm) between the tip and local surface topography. By using the stored topographical
data instead of the standard feedback, the separation remains constant without sensing the surface. At
this height, cantilever amplitudes are sensitive to electric field force gradients or relatively weak but long-
range magnetic forces without being influenced by topographic features. Two-pass measurements are
taken for every scan line, producing separate topographic and magnetic force images.
1.3.2.5.2 Electrochemical AFM (ECAFM)
This option allows us to perform electrochemical reactions on the AFM. It includes a potentiostat, a fluid
cell with a transparent cantilever holder and electrodes, and the software required to operate the poten-
tiostat and display the results of the electrochemical reaction.
1.3.2.5.3 Stand-Alone AFM
Digital Instruments, Inc., also manufactures a stand-alone AFM which measures the topography of a
sample with subnanometer resolution regardless of the size of the sample (Anonymous, 1991). The stand-
alone AFM can be placed directly on large samples (larger than about 10 × 10 mm) which cannot be
fitted into the AFM assembly, Figure 1.22. Either the sample must be larger in diameter than the three
support posts or the sample must be rigidly mounted to a larger substrate. Scan lengths of this instrument
are 75 and 125 µm. In these units, the sample is stationary. The cantilever beam and the compact assembly
of laser source and detector are attached to the free end of a piezoelectric transducer, which drives the
tip over the stationary sample, Figure 1.22A and B. Because the cantilever beam and detector assembly
are scanned instead of the sample, some vibration is introduced and lateral resolution of this instrument
is reduced. In the stand-alone AFMs, a single photodetector instead of split photodiode detector is used.
As a result, friction force measurement (to be discussed later) cannot be made.
A cylindrical piezoelectric tube scans a very sharp tip which is mounted on a flexible cantilever over
the sample surface. A compact interferometric detection system mounted on the end of the piezotube
senses the deflection of the cantilever as features in the sample are encountered. In the most common
operating mode, the control system varies the Z-voltage applied to the piezo to keep the cantilever
deflection nearly constant as the tip is scanned over the sample surface in a raster pattern. The variation
in the Z-voltage applied to the piezo translates directly into the variation in height across the sample.
The interference system used to detect cantilever deflection can be made quite compact and therefore
is mounted directly on the piezotube. Figure 1.22C shows the cantilever deflection detection system. The
laser diode emits light from both the top, beam 2, and the bottom, beam 1. The light emitted from the
bottom of the laser is reflected off the cantilever and back into the laser. The reflective cantilever forms