
5.1 Classic Fluorescence Lifetime Experiments 79
5.1.8 Fluorescence Anisotropy
If the fluorescence of a sample is excited by linearly polarised light the fluores-
cence is partially polarised (see Fig. 5.7 and Fig. 5.8). The fluorescence anisot-
ropy, r, is defined as
)(
)()(
)(
tI
tr
sp
(5.7)
with I
p
= fluorescence intensity parallel to excitation, I
s
= fluorescence intensity
perpendicular to excitation, I = total intensity.
The fluorescence anisotropy decays with the rotational relaxation time,
W
rot
. The
relaxation time is an indicator of the size of the dissolved molecules, dye-solvent
interactions, aggregation states, and the binding state to proteins [102, 308, 549].
Typical rotational relaxation times of free fluorophores in water are in the range
from 50 to 500 ps.
Time-resolved measurements of anisotropy are difficult because I
p
(t) - I
s
(t) is
small compared to the fluorescence components I
p
(t) and I
s
(t) themselves.
I
p
(t) and I
s
(t) are detected with different efficiency, especially if a monochromator
is used. The effect depends on the angle of the grating, i.e. on the wavelength, and
on the slit width and the beam geometry. Anisotropy measurements therefore
require calibration of the efficiency of the I
p
(t) and I
s
(t) detection channels. The
relative efficiency, E
p
/E
s
, of the I
p
and I
s
detection channels is the „G factor“:
p
E
E
G
(5.8)
There are two ways to determine the G factor [308, 389]. The first one is to run
a measurement with horizontal polarisation of the excitation beam. For an angle of
90° between the optical axis of excitation and emission, the excited-state distribu-
tion is oriented towards the axis of observation. Consequently both channels
measure equal perpendicular components. The ratio of the measured intensities
represents the G factor.
The second way to obtain G is „tail matching“. A sample with a depolarisation
time substantially shorter than the fluorescence lifetime is measured. The G factor
is obtained from the intensities in the later parts of the decay curves. The advan-
tage of tail matching is that it can be used also for optical systems with excitation
and detection along the same optical axis.
To separate the rotational relaxation from the fluorescence decay it is essential
that the correct total intensity, I(t) is taken for the denominator of r(t). It is nor-
mally assumed that both the excitation and emission light cones have negligible
numerical aperture, and that the excitation is polarised perpendicularly to the plane
defined by both optical axes, see Fig. 5.8.
The total intensity is then I(t) = I
P
(t) + 2 I
S
(t). The factor of 2 results from the
geometrical fact that light polarised longitudinally to the optical axis of observa-
tion is not detected [308, 549]. The situation for a microscope lens of high NA is
different. Both focusing and detection under high NA results in the conversion of
transversal E vectors into longitudinal ones, and vice versa [18, 426, 465, 466,
551], see Fig. 5.9. The total intensity is therefore I (t) = I
P
(t) + k I
S
(t), with k < 2.