
14
MEASUREMENT
AND
DETECTION
OF
RADIATION
The signal that comes out of the detector is very weak, in the millivolt (mV)
range (Fig. 1.3). Before it can be recorded, it will have to be amplified by a
factor of a thousand or more. To achieve this, the signal will have to be
transmitted through a cable to the next instrument of the counting system,
which is the amplifier. Transmission of any signal through a cable attenuates it
to a certain extent. If it is weak at the output of the detector, it might be lost in
the electronic noise 'that accompanies the transmission. This is avoided by
placing the preamplifier as close to the detector as possible. The preamplifier
shapes the signal and reduces its attenuation by matching the impedance of the
detector with that of the amplifier. After going through the preamplifier, the
signal may be safely transmitted to the amplifier, which may be located at a
considerable distance away. Although some preamplifiers amplify the signal
slightly, their primary function is that of providing electronic matching between
the output of the detector and the input of the amplifier.
There are many types of commercial preamplifiers, two of which are shown
in Fig. 1.8. In most cases, the HV is fed to the detector through the preamplifier.
1.5.6
The
Amplifier
The main amplification unit is the amplifier. It increases the signal by as many
as 1000 times or more. Modern commercial amplifiers produce a maximum
signal of 10 V, regardless of the input and the amplification. For example,
consider a preamplifier that gives at its output three pulses with heights 50
mV,
100 mV, and 150 mV. Assume that the amplifier is set to 100. At the output of
the unit, the three pulses will be
Note that the third value should be 15
V,
but since the amplifier produces a
maximum signal of 10 V, the three different input pulses will show, erroneously,
as two different pulses at the output. If only the number of particles is
measured, there is no error introduced-but if the energy of the particles is
measured, then the error is very serious. In the example given above, if gammas
of three different energies produce the pulses at the output of the preamplifier,
the pulses at the output of the amplifier will be attributed erroneously to
gammas of two different energies. To avoid such an error, an observer should
follow this rule:
Before any measurement of particle energy, make certain that the highest pulse of the
spectrum to be measured is less than
10
V
at the output of the amplifier.
In addition to signal amplification, an equally important function of the
amplifier is to convert the signal at the output of the preamplifier into a form