276 Radiation Dosimetry: Instrumentation and Methods
coefficients of the detector and the medium, and d is a
weighting factor which takes into account the size of the
detector. It is given by
(4.45)
where
is the effective mass attenuation coefficient for
electrons and g is the average path length through the
detector. Burlin originally used the following expression
for
g:
,
where V is the volume and S is the total surface area.
The ratios of mass energy absorption coefficients for
different TLDs and water, weighted over energy spectra
given by Seltzer [52], for scattered components and for
the whole spectrum are shown in Figure 4.73
The experimental results of dose distribution measure-
ments with different TL dosimeters are shown in Figure 4.74.
Doses in water from the results of the Fricke dosimeter
were calculated.
EGS4 Monte Carlo simulations have been performed
by Mobit et al. [53] to examine general cavity theory for a
number of TLD cavity materials irradiated in megavoltage
photon and electron beams. The TLD materials were LiF,
Li
2
B
4
O
7
, CaF
2
, and CaSO
4
irradiated in Perspex, water, Al,
Cu, and Pb phantoms. For megavoltage photon beams, this
has been done by determining the dose component (1
d)
resulting from photon interactions in the cavity compared
with the dose component resulting from photon interactions
in the phantom material (d) by Monte Carlo simulations
and analytical techniques. The results indicate that the Burlin
exponential attenuation technique can overestimate the dose
contribution from photon interactions in a 1-mm thick LiF
cavity by up to 100%, compared with the Monte Carlo
results for LiF TLDs irradiated in a water or Perspex phan-
tom. However, there is agreement to within 1% in the quality
dependence factor, determined from Burlin’s cavity theory,
Monte Carlo simulations, and experimental measurements
for LiF and Li
2
B
4
O
7
TLDs irradiated in a Perspex or a water
phantom. The agreement is within 3% for CaF
2
TLDs.
However, there is disagreement between Monte Carlo sim-
ulations and Burlin’s theory of 6 and 12% for LiF TLDs
irradiated in copper and lead phantoms, respectively. The
adaptation of Burlin’s photon cavity theory and other mod-
ifications to his photon general cavity theory for electrons
have been shown to be seriously flawed. [53]
Mobit et al. presented cavity theory in the following
way. When a detector is placed in a medium exposed to
ionizing radiation in order to measure dose, it forms a
cavity in that medium. The cavity is generally of different
atomic number and density to the medium. Cavity theory
gives the relation between the dose absorbed in a medium
(D
med
) and the average absorbed dose in the cavity decay.
(4.46)
where f
med,cav
is a factor that varies with energy, radiation
type, medium, size, and composition of the cavity. For a
cavity that is small compared with the range of the electrons
incident on it in a photon beam, the Bragg-Gray relation
applies and f
med,cav
is given by the average mass collision
stopping power ratio of the medium to the cavity (s
med,cav
),
generally evaluated according to Spencer and Attix. For a
cavity that is large compared with the electrons incident on
it, f
med,cav
is given by the ratio of the average mass energy
absorption coefficient of the medium to the cavity material.
The Burlin cavity theory equation (4.44) requires a
value to be provided for d which represents the proportion
FIGURE 4.72 Depth-dose curves generated from normalized and scaled Y-90 beta particle spectrum source for varying energy ranges
in LiF. (From Reference [50]. With permission.)
d 1 e
g
()
g
4 VS()
D
med
D
cav
f
med,cav
Ch-04.fm Page 276 Friday, November 10, 2000 12:01 PM