The presence of high-order lines (Section 5.3.1) is a complicating factor in
WD spectra. They can be suppressed by means of pulse-height analysis,
thoughnotalwayscompletely(seeSection5.3.4.1).
7.4 Mineral identification
The main object of finding out which elements are present is usually to identify
the phase. Common rock-forming minerals contain principally the following
major elements: Na, Mg, Al, Si, S, P, K, Ca, Ti, Cr, Mn and Fe (also O, which
is detectable only with a thin-window ED detector, however). For sulphides
and related phases additional elements are relevant, e.g. Co, Ni, Cu, Zn, Pb
and sometimes other elements such as As, Se, Sb and Bi. For carbonates C, like
O, is detectable only with a thin-window detector.
Examples of ED spectra of common minerals are given in the appendix.
Different minerals containing the same major elements in different pro-
portions are often distinguishable on the basis of the relative heights of the
peaks (e.g. for pyroxenes the Si peak is larger relative to Fe and Mg than for
olivines).
Mineral identification based on electron microprobe data is facilitated by
using a database such as that described by Smith and Leibowitz (1986).
Automatic mineral identification is useful in modal analysis (Section 6.10).
For limited groups of minerals in specific types of rocks quite simple methods
of mineral identification are adequate. For example, Nicholls and Stout (1986)
ranked major elements in silicates in order of peak intensity as the basis for
identification, supplemented by intensity ratios when necessary.
7.5 Quantitative WD analysis
The electron microprobe is designed for quantitative analysis as its principal
function, for which WD spectrometers are normally used, as described in the
present section. An SEM equipped with a WD spectrometer is less satisfactory
for this purpose in some respects. More commonly SEMs have only an ED
spectrometer, which can be used quantitatively, with some restrictions, as
discussed in the next section.
When using WD spectrometers, a suitable crystal must be chosen for each
line (Section 5.3.1) and appropriate conditions for pulse-height analysis
selected (Section 5.3.4). The measured intensities need to be corrected for
dead-time (Section 5.3.5). The beam current should be chosen to be of such a
value as to avoid count-rates above about 5 10
5
counts per second, which
could cause pulse-height depression and unduly large dead-time corrections. For
7.5 Quantit ative WD analysis 115