Infrared Absorption Spectroscopy 123
of wavelength. This approach also no longer requires dithering the laser wavelength or
the cavity by a piezoelectric transducer, as required by CEAS and ICOS, respectively, to
effectively couple the laser into a large number of transverse cavity modes.
Both OA-ICOS and CEAS have been most successfully applied in the near-IR spectral
region, where Baer et al. (2002) report A
min
values of 3 ×10
−11
cm
−1
Hz
−1/2
. Such high
performance, which is approximately one order of magnitude better than our field
second harmonic measurements in the mid-IR, needs to be further verified by repetitive
measurements. In the near-IR spectral region, these approaches take advantage of the
convenience of commercially available high power (10–50 mW) tunable telecom diode
lasers, low noise detectors D
∗
∼10
12
cm Hz
1/2
W
−1
, and high quality reflective mirror
coatings > 9999%. Translation of this performance to the mid-IR spectral region is
limited by: (1) higher noise detectors D
∗
∼10
11
cm Hz
1/2
W
−1
; (2) adequate tunable
laser sources of sufficient output power (tens of mW) and high beam quality; and
(3) similar-performing highly reflective coated mirrors. However, if one could ultimately
circumvent these issues and translate this spectroscopic performance to the mid-IR
spectral region, one could achieve extremely low mixing ratio detection limits for a
whole series of atmospherically important gases. As discussed previously, the fundamental
vibrational-rotational transitions are several orders of magnitude stronger in the mid-
IR than the near-IR. For example, spectral line intensities for CO
2
in the mid-IR at
43 m are ∼20 000 times stronger than near-IR transitions at 16 m where OA-ICOS
has successfully been employed. In such instances, medium pathlength-normalized A
min
values of 10
−8
cm
−1
A
min
= 10
−4
l= 100m in the mid-IR would require an equivalent
A
min
of 5 ×10
−13
cm
−1
to achieve the same sensitivity in terms of number density in the
near-IR. Despite this disadvantage, near-IR OA-ICOS achieves respectable mixing ratio
detection limits for the more abundant atmospheric gases (Baer et al., 2002).
2.5.1.3.6 BACKGROUND STRUCTURE AND
MEASUREMENT PERFORMANCE
Most, if not all, IR spectrometers employing solid-state laser sources are ultimately limited
by the stability of the background structure on which the absorption resides. This is the
case for direct absorption, WMS and FMS approaches just presented. An example of
such background structure is displayed in Figure 2.16 along with ambient, calibration,
and fit absorption profiles for CH
2
O. The background structure, which is due to optical
noise, plays a significant role in the determination of the ultimate mixing ratio one
deduces. This is particularly true here where the ambient CH
2
O mixing ratios are near the
detection limit. In this case, our airborne CH
2
O results would be in error by ∼100pptv
if the background spectrum is not first removed. Figure 2.17 shows this even more
dramatically in the case of our laboratory CH
2
O measurements where an input-mixing
ratio of 3.714 ppbv was sampled. Backgrounds were acquired for 10 seconds before and
after each 20 second sample, and including the sample, background, and intervening
flushing periods, each data point represents a 1 minute acquisition. In the upper panel
the backgrounds were ignored, while the average of the two backgrounds was subtracted
before fitting in the lower panel. In both cases, the sample data were fit to a CH
2
O
standard acquired previously using a real-time multiple linear regression approach (Sewell
et al., 1993). The retrieved CH
2
O mixing ratio averaged 4036 ±0305 ppbv in the case
of the upper panel and 3714 ±0055 ppbv using background subtraction in the lower