290 Part B Chemical and Microstructural Analysis
seven repeat measures of one peak followed by seven of
the other allows the trend in the ratio during acquisition
to be evaluated as well as the intensity ratio repeatability
standard deviation. With around 2 M counts per channel
at the peaks, ISO 24236 shows how repeatability stan-
dard deviation of better than 0.5% may be attained if the
data are recorded at 0.1 eV energy intervals and specific
Savitzky and Golay smoothing [6.37] is used. Any drift
in the absolute intensities or of the ratio between mea-
surements may indicate a source, analyzer or detector
instability that then needs to be investigated.
Calibrating the Intensity Scale
Interlaboratory studies to compare the shapes of spec-
tra obtained in different laboratories unfortunately show
that there are marked differences [6.61] that can lead to
variations of a factor of two in quantification if the same
relative sensitivity factors are to be used in all labora-
tories. These differences exist between similar models
of spectrometer from the same manufacturer and arise
mainly from the age dependence of the detector effi-
ciency D(E). D(E) exhibits a curve that rises with the
detected electron kinetic energy E from zero at E =0
to a maximum in the energy range 200–600 eV and then
to a slow decline at higher energies. In addition to the
detector efficiency, there are electron optical terms to
describe the spectrometer transmission function T(E).
These need to be combined to give the total instrumental
response.
Formally, one may write the intensity–energy re-
sponse function (IERF)as
IERF = T (E) D(E) , (6.5)
with additional terms, omitted here, that may arise from
stray electron or magnetic fields [6.62]. The term T (E)
is usually approximately proportional to E for spec-
trometers operated in the constant ΔE/E mode, and
proportional to E
−n
, where n ranges from 0 to 1, as
the energy increases in the constant ΔE mode. In the
constant ΔE/E mode, all voltages on electron optical
elements of the spectrometer are scanned so that they
remain in fixed proportion to each other. The resolu-
tion then deteriorates as the energy increases. This is the
mode generally used for AES, unless high-resolution
spectra are required, since very simple spectrometers
may then be used with high efficiency and with high in-
tensities at the high energies where the peaks are weak.
On the other hand, the constant ΔE mode is used for
high-resolution analysis so that ΔE, the spectrometer
energy resolution, is maintained at, say, 0.25 eV at all
energies. This is usually achieved by setting the pass
element of the spectrometer to detect, say, 25 eV elec-
trons, and then scanning this through the spectrum. If
we know the IERF, the true spectrum that we need,
n(E), is given by
n(E) =
I(E)
IERF
, (6.6)
where I(E) is the measured spectrum.
In order to calibrate spectrometers for their absolute
or relative IERFs, a series of studies were made using
different configurations of an instrumented spectrom-
eter with a Faraday cup detector to measure absolute
reference spectra [6.61–64]. These spectra were meas-
ured for Cu, Ag, and Au polycrystalline foil samples
using a 5 keV electron beam at 30
◦
to the surface
normal. Using these spectra, the absolute IERF may
be determined for any spectrometer. To facilitate this,
a software system has been designed for users to
self-calibrate their instruments based on their own mea-
surements for these foils [6.65]. The reason for using
three foils when, in principle, one would suffice, is
to evaluate the scatter between the three independent
IERF derivations in order to calculate the repeatability
of the average IERF derivation. These derivations can
be consistent to < 1%. In the calibration, certain other
diagnostics are important. For instance, internal scatter-
ing [6.66] may occur in some spectrometers, and if this
has any significant intensity, it leads to uncertainty in
the derived IERF. The above-mentioned software diag-
noses the extent of the internal scattering using the rules
established in [6.66] with the Cu and Ag samples.
The true spectral shape obtained in this way will
not change significantly with time provided the IERF
is determined at appropriate time intervals. Being ab-
solute, use may then be made of an extremely large
volume of theoretical knowledge as well as background
removal procedures based on physically meaningful al-
gorithms [6.67,68] in order to interpret different aspects
of the spectra.
As noted earlier, many analysts do not use any
significant theoretical evaluation of the spectra and sim-
ply use the peak-to-peak differential intensity. Relative
sensitivity factors for (6.1) are available from several
handbooks [6.39–43], but analysis shows that the lack
of control of the IERF leads to significant variability.
Additionally, different choices of modulation energy for
the differentiation increase that variability from source
to source [6.69], so that half of the published sensitiv-
ity factors for each element differ from the average by
more than a factor of 1.5. These issues are addressed
below.
Part B 6.1