The search unit was centered with respect to the reference tube, ensuring that a water gap between the search unit and the
tube was equal to the focal length of the search unit minus the radius of the tube. Angular adjustment and positioning
were performed to obtain a maximum response from the tube wall indicative of proper centering. For circumferential
inspection to detect longitudinal flaws, the search unit was displaced laterally or set off from the center of the tube to
obtain the proper angle at which the sound would travel through the tube wall (inset, Fig. 6). Generally, inspection was
performed using a shear wave traveling at an angle of 45° around the tube. The offset distance required was calculated
using the formula:
L = (D/2)(v
w
v
m
) sin θ
(Eq 1)
where L is the offset distance, D is the diameter of the tube or pipe, v
w
is the velocity of sound in water, v
m
is the velocity
of sound in the tube, and is the desired refracted angle in the tube or pipe. The value for the velocity of sound in the tube
depended on whether a longitudinal or shear wave was to be used. The offset value obtained from Eq 1 was an
approximation. The actual amount of offset was adjusted from that calculated to obtain best presentation and equalization
of responses from outside and inside notches.
When the search unit and reference tube or pipe had been properly located with respect to each other, the reference tube
was rotated and driven longitudinally past the search unit. Controls of the ultrasonic instrument were adjusted to display a
clear response from both inside and outside notches on an oscilloscope screen. The controls of the flaw-alarm module of
the instrument were adjusted to position the gate properly to include the signals from the notches and to activate the alarm
when they were detected. The pulse-repetition rate of the instrument was adjusted high enough to ensure the detection of
all notable flaws at the speed of inspection.
The inspection speed was controlled by the rate of rotation and longitudinal movement per revolution (pitch) of the tube.
The allowable pitch was a function of the length of the line-focused ultrasonic beam and the size of the flaw to be
detected. Normally, when the reference standard was passed through the unit, the controls had been adjusted to provide
more than one signal from each of the notches. When this had been established, the system was properly calibrated for
production inspection.
Ultrasonic Inspection Precautions. Generally, a chart recorder is employed to provide a permanent record of the
inspection described in Example 2. Multiple search units can be used in the immersion tank to provide several
simultaneous inspections during one pass of the pipe or tube through the unit. Specifications may call for circumferential
inspection from two directions because the reflection from a flaw may vary, depending on the direction in which the
ultrasonic beam strikes it. In addition to the circumferential inspection to detect longitudinal flaws, a longitudinal (axial)
inspection may be required to detect transverse flaws. Also, it may be desirable to ultrasonically measure wall thickness,
eccentricity, or both. All these tests can be performed at the same time by utilizing search units that are designed for the
tests and that are properly positioned in the tank.
Normally, rejection is based on the presence of flaw indications exceeding those from the reference notch. Reworking and
reinspection are generally permitted if other requirements, such as minimum wall thickness, are satisfied.
Other refinements are included in, or can be added to, the inspection system. For example, the feeding of tubes to the unit
and their withdrawal can be automated. Various audible and visible alarm systems and marking devices can be added.
Normally, the water used in the system is filtered and deaerated. Air entrapped in the water can produce false indications.
Similarly, water on the inside surfaces of tubes will produce false signals, and these surfaces must be kept dry. Tubes are
connected to each other by stoppers or by taping the ends together. The glands at each end of the water tank must be cut
in a manner that will allow passage of the tubes without undue loss of water couplant. Finally, air must be prevented from
being drawn into the entry gland along with the tube. This is usually accomplished by directing a stream of water over the
outside of the tube just before it enters the gland.
In eddy current inspection, use of an encircling detector coil is limited to a maximum tube diameter of about 75 mm
(3 in.). As tube diameter increases, the ratio of flaw size to tube diameter decreases; consequently, the flaw is increasingly
more difficult to detect. This problem is overcome by using several small probe coils (Ref 6, 7, and 17) and with spinning
probes (Ref 8). When probe coils are used, the flaw becomes a significantly high percentage of the zone surveyed.
Because independently mounted probes ride over the tube surface, good magnetic coupling is ensured. Magnetic
saturation is used to obtain maximum sensitivity to flaws close to, or on, the inside surface of the tube, and the frequency