
PHOTON (GAMMA-RAY AND X-RAY) SPECTROSCOPY
411
spectra. Equation 12.10, developed for coaxial detectors, predicts the efficiency
with an accuracy of 0.2 percent over the energy range 160-1333 keV. Further
testing of 12.10 showed2' that the last term involving the constants
a,
and
a,
has a negligible effect on the result.
Equations 12.8-12.10 are given just as examples that indicate the general
energy dependence of the efficiency curve. What is done in practice is to
measure the efficiency as a function of energy using calibrated sources emitting
gammas of known energies and
inten~ities."-'~.~'
A
table of gamma energies
used for calibration is given in App. C.
12.7.2
Energy Resolution of Ge Detectors
The energy resolution of a Ge detector is given in terms of the FWHM
(r).
The
width
r
consists of the following two components:
r,
=
width due to detector effects
re
=
width due to effects of electronics
Since these two components are uncorrelated, they are added in quadrature to
give the total width,
T,
r=,/m
(12.11)
As shown in Chap.
9,
the width
T,
is energy dependent and is given by
where
F
is the Fano factor and
w
is the average energy needed to produce an
electron-hole pair. For germanium, at the operational temperature of
77
K,
w
=
2.97 eV. Thus,
The width
re
increases when the detector capacitance increases. The
detector capacitance, in turn, generally increases with detector size and may
change with detector bias. Good Ge detectors have a flat capacitance-bias
relationship over most of the range of bias voltage applied.
The capacitance of the detector has an effect on the energy resolution
because it influences the performance of the charge-sensitive preamplifier that
accepts the detector signal. The contribution of the preamplifier to the value of
re
increases with the input capacitance. One of the manufacturers, Canberra,
reports a 0.570-eV
re
with zero input capacitance and a slow increase with
higher values as shown in Fig. 12.32. Clearly, the resolution improves if the
capacitance is kept low. The other component of the input capacitance comes
from items like connectors and cables. Reduction of the length of input cable
and of connectors' capacitance is helpful. For the best resolution with a given
system, the preamplifier should be located as close to the detector as possible.