
ELECTRONICS
327
ready for the next pulse. Without pulse shaping, the detector signal lasts so
long that pulses overlap. If only the number of particles is counted, pulse
overlap leads to loss of counts (dead time loss). In spectroscopy measure-
ments, pulse overlap worsens the resolution.
2.
To improve the signal-to-noise ratio.
Noise created in the detector and the
early amplification stages accompanies the detector signal. Appropriate pulse
shaping can enhance the signal while at the same time reduce the noise.
Thus, the signal-to-noise ratio will improve, which in turn, leads to better
energy resolution.
3.
For special pulse manipulation.
The detector pulse may, in certain applica-
tions, need special pulse shaping to satisfy the needs of certain units of the
counting system. As an example, the signal at the output of the amplifier
needs to be stretched before it is recorded in the memory of a multichannel
analyzer (see Sec. 10.12).
The pulse-shaping methods used today are based on combinations of
RC
circuits and delay lines. For example, the use of a
CR-RC
circuit combination
produces the pulse shown in Fig. 10.15. The exact shape and size of the output
pulse depends on the relative magnitudes of the time constants
C,
R,
and
C, R,.
The use of the
CR-RC
circuit combination provides, in addition to pulse
shaping, a better signal-to-noise ratio by acting as high-pass and low-pass filter
for undesired frequencies.
If one adds more
RC
integrating circuits, the pulse will approach a Gauss-
ian shape (Fig. 10.16).
If one applies a
CR-RC-CR
combination, the result is a doubly differenti-
ated pulse as shown in Fig. 10.17. Commercial amplifiers usually provide either
singly or doubly differentiated pulses. In all cases, the final pulse is the result of
repeated application of
CR-CR
circuits. Figure 10.18 shows such pulses pro-
duced by the application of many
RC
and
CR
circuits, called unipolar and
bipolar pulses, respectively. Pulse shaping using a delay line and an
RC
circuit is
shown in Fig. 10.19.
Input signal
A
~utpu; signal
Figure
10.15
An
example
of
CR-RC
shaping. The triangle indicates the amplification unit
(A)
that
isolates the
two
shaping circuits.