
H.8 Example- design of an optimum compressor cascade
299
command before returning to CASCADE. Then press the Print Screen key whenever
you wish to dump the graphical presentations to the printer.
11.8 Example- design of an optimum compressor cascade
To illustrate the use of the program CASCADE let us consider the selection of
optimum profile geometry to meet prescribed aerodynamic requirements which are
to be as follows:
Design requirements
We will adopt the case considered in Chapter 2, Fig. 2.7, where we are given
fll ----
54.59~
f12 =
30.69~ C4 profile, circular arc camber-line. Let us also specify that
the loading must be conservative compared with the maximum allowable diffusion
factor of DF = 0.6.
The first task is to estimate a suitable value for the pitch/chord ratio
t/l.
Substituting
the above data into Eqn (2.30) for DF = 0.6 gives the maximum allowable value
t/l
= 1.1623. Since we are asked to produce a conservative design let us adopt the
slightly smaller value
t/l
= 1.0. Back-substitution into Eqn (2.28) then yields a
conservative diffusion factor of DF = 0.5618.
Now we can proceed to use the program CASCADE to find the camber 0 and
stagger A values to deliver the required
~2
with shock-free inflow/31. There are two
practical approaches as follows.
Solution - first method, interpolation
For a first crude estimate of cascade geometry we could guess that the stagger might
approximate to the average of
fll
and 132 (see Fig. 2.5 for a perception of this). On
the other hand we might expect the camber angle to be rather bigger than the fluid
deflection, say 1.5 x (131-/32). Thus our first estimation will be as follows:
A ~ 1(54.59~ + 30.69 ~ = 42.64 ~ 43 ~
0 ~- 1.5 • (54.59 ~ 30.69 ~ = 35.85~ ~ 36 ~
The strategy to be followed now is to study four test cascades around these values
using the program CASCADE in order to predict their shock-free performance, and
then to interpolate to estimate the required stagger and camber. The four test cases
chosen for illustration are shown in Table II.1.
Table II.1 Test cases for program CASCADE
Shock-free inlet and outlet
angles from CASCADE program
Case no. A ~ O~
~opt ~opt
1 40 35 52.37 31.32
2 40 45 55.22 28.23
3 45 35 57.09 36.82
4 45 45 59.79 33.80