pass through all the GI sphincters with relative ease, including
the lower oesophageal sphincter and the pyloric sphincter. The
capsulemustalsobecheaptomakesinceitwillonlybeused
once. Low power consumption is a requirement to minimise
battery, hence overall, size and increase operating time. Gastric
emptyingtakesbetween30minand4htocompleteandsemi-
digested food (chyme) takes 2–3 h to pass through the small
intestine. Once in the colon, gut content moves relatively slowly
at approximately 5–10 cm/h. Peristaltic waves in the colon are
known as mass movements and only occur 1–3 times per day.
Overall the capsule might take a maximum of 8 h to traverse the
upper alimentary tract and the small intestine, while a complete
passage through the GI tract might take up to 32 h. Using
readily available silver oxide battery technology, an energy sto-
rage density of 500 mWh/mL can be achieved (30), thus a
suitable source, such as two SR48 cells (110 mWh each) could
deliver enough energy to complete small intestinal measure-
mentsifthepowerconsumptionwaslessthan20mW.
The data sampled in the GI tract by a capsule must be
retrieved accurately and securely. This usually means that the
data must be wirelessly transmitted and correctly received by a
device worn by the patient. There are a number of radio com-
munication standards encompassing several international
industrial, scientific and medical telemetry bands (pan-Eur-
opean medical device frequency allocations (31), and the US
Federal Communications Commission frequency allocations
for biomedical telemetry and industrial, scientific and medical
(ISM) devices – regulations S5.150, US209 and US350). The
main bands of interest are at 418 MHz, 434 MHz, 868 MHz
and 915 MHz.
As with all measuring devices, the user must be confident that
the data retrieved is accurate. The problems of accuracy can be
dealt with via the normal techniques of instrument design and
calibration. However, an additional constraint for wireless devices
is that the data must be secure. Since capsule devices operate in the
unlicensed ISM frequency bands there is a severe risk of interfer-
ence that could be particularly dangerous in the context of a
medical device. Of necessity, the sensors and signal acquisition
electronics require analogue circuits. In the early devices the entire
design was analogue, making the data transfer from the devices
extremely insecure. However, modern electronic techniques per-
mit designers to convert the analogue signal to the digital domain
within the capsule, enabling the use of secure digital wireless
techniques. These techniques ensure that data from any given
capsule can be uniquely identified to avoid attributing diagnostic
information to the wrong individual. Details of such designs, and
others concerning wireless sensor systems, may be found in the
literature (32, 33).
Wireless Endoscopy: Technology and Design 229