Field Measurements of Atmospheric Composition 49
is a mixture of hydrocarbons for the calibration of gas chromatographs that are able to
measure many species during a single elution through the column (Chapter 8). If the
concentrations provided are higher than ambient levels, then a dilution system must be
designed, and mass flow-controllers should be calibrated carefully. It is often prudent to
use in-line scrubbing agents to remove unwanted impurities, even if the manufacturer
does not list them, as degradation in the cylinder can occur over time. An example is the
removal of other oxides of nitrogen from a cylinder of nitric oxide (NO).
Unfortunately, some species are unstable, and decompose shortly after preparation,
either in the gas-phase or are catalysed by surfaces. Examples include oxygenated VOCs
and HNO
3
. In some cases, it is possible to use permeation tubes containing liquid samples
held at a known temperature that generate a gaseous sample at a known rate into a flow of
air. The permeation tube can be weighed at intervals to establish the amount of material
released. Examples are formaldehyde, HNO
3
,Cl
2
and HCHO, for which commercial
devices are available. For others, it is not possible to purchase the trace gas at all, and it is
must be prepared carefully in the laboratory followed by purification. Examples include
ozone, which can be prepared using a silent discharge in O
2
, or UV photolysis of O
2
(typically using the 184.9 nm line of a mercury lamp), and nitrous acid, HONO, which
can be prepared using several methods, usually involving the reaction of a strong acid
with a salt containing the nitrite ion, NO
2
−
.
The calibration of instruments to measure free-radical concentrations in the atmosphere
represents a difficult challenge, as free-radicals are very short-lived and are removed
rapidly at surfaces. No calibration is necessary for free-radical measurements by open-
path differential optical absorption spectroscopy or cavity ring-down spectroscopy (e.g.
NO
3
, IO, see Chapter 3), as the methods are self-calibrating, but absorption methods
that use enclosed multipass cells still need to be calibrated to assess any wall losses. It
is necessary to generate the free-radicals at the total atmospheric pressure sampled by
the instrument (ideally in air). Methods normally involve a chemical titration reaction,
or the photolysis of a suitable precursor molecule, sometimes followed by conversion
to the radical of interest. For OH radicals, a known concentration in a flow at a total
pressure of a few Torr (suitable for in situ instruments that measure in the stratosphere)
can be generated from the reaction H +NO
2
→OH +NO, with H atoms generated in a
microwave discharge of H
2
.NO
2
is the limiting reagent, and its partial pressure in a flow
is calculated from the ratio of the flow of NO
2
(measured using a calibrated mass flow
controller) to the total flow (NO
2
+H
2
+ any buffer gas). This method is not suitable
for the calibration of instruments to measure OH in the troposphere at close to one
atmosphere total pressure. In this case, the most commonly used method is the photolysis
of water vapour at 184.9 nm, using radiation from a mercury lamp, in a flow of synthetic
air which generates OH and HO
2
radicals in a 1:1 ratio. If only a source of HO
2
radicals
is required, addition of CO to the calibration gas mixture rapidly converts OH to HO
2
radicals. For CH
3
O
2
, the photolysis of CH
3
I at 254 nm followed by the fast reaction of
CH
3
with O
2
has been used to generate CH
3
O
2
radicals. Using OH as an example, the
OH concentration following the photolysis of water vapour is given by:
OH = H
2
O
H
2
O1849nm
OH
F
1849nm
t (1.6)
where is the water vapour absorption cross-section, is the photodissociation quantum
yield of OH and F is the photon flux of the lamp, all at 184.9 nm, and t is the photolysis