7.2.2. Transmission Electron Microscopy (TEM)
A transmission electron microscope is analogous to a slide projector, with illumina-
tion from an electron beam rather than light. When an electron beam is imping ed
upon a sam ple, a black and white TEM image is formed from the passage of some
electrons through the sample untouched, alon gside the combination of interactions
between other electrons and sample atoms (e.g., inelastic/elastic scattering, diffrac-
tion). If the undiffracted beam is selected to form the image, it is referred to as
bright-field imaging; in contrast, selection of strongly diffracting regions of the
sample, which would appear brighter than the transmitted beam, is known as
dark-field imaging.
[19]
It should be noted that electrons may also be absorbed by
molecules containing large atoms, or by surface contamination (e.g., dust, grease).
The absorption of a high density of electrons in a specific region will cause a buildup
of heat, leading to sample destruc tion and poor image quality.
Analogous to throwing a baseball of varying speeds through a wall, the relative
degree of penetration through a particular sample is governed by the energy of the
electron source. That is, higher energy electrons (e.g., 200 keV vs. 100 keV) will be
more penetrating, allowing for the characterization of thicker and/or less transparent
samples. In general, increasing the thickness of a sample, or decreasing the energy
(i.e., accelerating voltage) of the electron beam, will induce more scattering events
through more effective interactions between the electron beam and atoms of the
sample. This effect will enhance image contrast, since there is a larger deviation
between the path lengths of transmitted and scattered electrons that reach the
viewing screen. However, this improvement of image quality is offset by plentiful
inelastic collisions that yield a broadened wavelength distribution of the electron
beam. Since individual electrons will have differing energies, they will be brought
into focus at different points resulting in a blurry image (i.e., decreased resolution).
Sample preparation techniques
Not unlike other materials (or molecular) characterization techniques, the most
important and time-consuming aspect of TEM analysis is sample preparation.
Specimens for TEM analysis are placed on special micromesh grids of a conductive
metal such as Cu, Au, or Ni. The typical dimensions of TEM grids are ca. 3 mm in
diameter and 10–25 mm thick. The mesh number of a grid indicates the number of
grid openings per linear inch. The smaller the grid number, the larger the hole size
and the greater the ratio of open area to covered area (Figure 7.9a). For example, a
200-mesh grid has 20 holes along its diameter; a 400-mesh grid has 40 holes.
This translates to hole sizes of ca. 200 mm for 100-mesh, ca.97mm for 200 mesh,
ca.63mm for 300 mesh, and ca.42mm for 400-mesh grids.
Most often, a grid is coated with a support film that holds the sample in place. The
film must be as transparent as possible, while providing support for the sample. To
prevent interference with electron-sample interactions, films containing light elements
(e.g., C, Be) are used. Support films are typically deposited onto the surface of grids
through a “floating” technique (Figure 7.9b). Amorphous carbon or polymeric
7.2. Electron Micros copy 597