84 Bharat Bhushan and Othmar Marti
Table 2.5. Noise in interferometers. F is the finesse of the cavity in the homodyne inter-
ferometer, P
i
the incident power, P
d
is the power on the detector, η is the sensitivity of the
photodetector and RIN is the relative intensity noise of the laser. P
R
and P
S
are the power in
the reference and sample beam in the heterodyne interferometer. P is the power in the No-
marski interferometer, δθ is the phase difference between the reference and the probe beam in
the Nomarski interferometer. B is the bandwidth, e is the electron charge, λ is the wavelength
of the laser, k the cantilever stiffness, ω
0
is the resonant frequency of the cantilever, Q is the
quality factor of the cantilever, T is the temperature, and δi is the variation in current i
Homodyne Heterodyne Nomarski
interferometer, interferometer interferometer
fiber-optic interferometer
Laser noise
δi
2
L
1
4
η
2
F
2
P
2
i
RIN η
2
P
2
R
+ P
2
S
RIN
1
16
η
2
P
2
δθ
Thermal noise
δi
2
T
16π
2
λ
2
η
2
F
2
P
2
i
4k
B
TBQ
ω
0
k
4π
2
λ
2
η
2
P
2
d
4k
B
TBQ
ω
0
k
π
2
λ
2
η
2
P
2
4k
B
TBQ
ω
0
k
Shot noise
δi
2
S
4eηP
d
B 2eη
(
P
R
+ P
S
)
B
1
2
eηPB
the other two. This means that certain light rays will propagate at different speeds
through the crystal than others. By choosing the correct polarization, one can se-
lect the ordinary ray or the extraordinary ray or one can get any mixture of the two
rays. A detailed description of birefringence can be found in textbooks(e.g., [150]).
A calcite crystal deflects the extraordinary ray at an angle of 6
◦
within the crystal.
Any separation can be set by choosing a suitable length for the calcite crystal.
The focus of one light ray is positioned near the free end of the cantilever while
the other is placed close to the clamped end. Both arms of the interferometer pass
through the same space, except for the distance between the calcite crystal and the
lever. The closer the calcite crystal is placed to the lever, the less influence distur-
bances like air currents have.
Sarid [116] has given values for the sensitivities of different interferometeric
detection systems. Table 2.5 presents a summary of his results.
Optical Lever
The most common cantilever deflection detection system is the optical lever [53,
111]. This method, depicted in Fig. 2.29, employs the same technique as light beam
deflection galvanometers. A fairly well collimated light beam is reflected off amir-
ror and projected to a receiving target. Any change in the angular position of the
mirror will change the position where the light ray hits the target. Galvanometers
use optical path lengths of several meters and scales projected onto the target wall
are also used to monitor changes in position.
In an AFM using the optical lever method, a photodiode segmented into two
(or four) closely spaced devices detects the orientation of the end of the cantilever.
Initially, the light ray is set to hit the photodiodesin the middle of the two subdiodes.
Any deflection of the cantilever will cause an imbalance of the number of photons