
32 3 Multidimensional TCSPC Techniques
tiphoton events. Because these events are not added into the photon distributions,
waveform distortions by pile-up are substantially reduced (please see also
Sect. 7.9.1, page 332).
Routing modules exist for different detector types. For detectors delivering
TTL output pulses, such as the Perkin Elmer SPCM-AQR APD modules or the
Hamamatsu H7421 PMT module, the router is relatively simple. The detector
pulses can be connected directly to the discriminators D1 to Dn.
In practice the noise sets a limit to the number of individual PMTs that can be
connected to a router. The cables connecting the PMTs to the router must be
matched with 50 Ohm resistors. Even with a near-perfect summing amplifier, the
noise from the matching resistors will be added to the output signal. Even worse is
noise from the environment picked up by the detectors. While resistor noise adds
quadratically, noise from the environment is more or less in phase for all detectors
and therefore adds linearly. In practice, no more than eight individual PMTs are
connected to one routing device.
A higher number of channels can be obtained if a multianode PMT is used. In a
multianode PMT the combined photon pulses of all channels are available at the
last dynode. This makes an external combination of the pulses unnecessary. The
noise problem is therefore lessened. Routers for multianode PMTs are combined
with the PMT tube into a common detector housing. TCSPC multichannel detec-
tor heads now exist for 16 channels. Using this method, devices with 32 channels
and even 64 channels appear feasible. In practice the number of channels is lim-
ited only by the power dissipation of the routing electronics.
It should be pointed out that the multidetector technique does not use any detec-
tor switching or multiplexing. Thus detectors need not be inactive for any fraction
of the acquisition time. Thus the multidetector technique can considerably im-
prove the counting efficiency of a TCSPC system. This is especially the case if a
fluorescence signal has to be recorded with spectral resolution. With a single de-
tector several measurements have to be performed one after another, usually by
scanning the spectrum by a monochromator. Most of the photons emitted by the
sample are then discarded. In a multidetector TCSPC system, on the other hand,
the signals of all wavelength intervals are detected simultaneously and loss of
photons is avoided.
Of course, the multidetector technique does not increase the maximum through-
put rate of a TCSPC system. In any TCSPC device there is a small but noticeable
loss of photons due to the „dead time“ of the processing electronics. The dead time
of advanced TCSPC devices is of the order of 100 ns, and for count rates above
1 MHz the counting loss becomes noticeable (see Sect 7.9, page 332). The counting
loss for a multidetector TCSPC system is the same as for a single detector system
operated at the total count rate of the detectors of a multidetector system.
An important and sometimes confusing feature of the multidetector technique is
that the relative counting loss is the same for all channels, independent of the
distribution of the rates over the detectors. The reason is that the photons detected
by all detectors are processed by the same TCSPC channel so that the counting
loss depends on the overall count rate. However, the photons appear randomly in
the particular detector channels. Therefore the dead time caused by a detection
event in one detector on average causes the same relative loss for all detector