nominal wall thickness plus the nominal height of weld reinforcement, not plus twice the nominal reinforcement as with
the double-wall, double-image technique. Radiation-source location with respect to the testpiece and the location of the
penetrameter on the block or shim are the same as for a double-wall, double-image technique.
In any setup for single-wall, single-image radiography, the penetrameters can be placed only on the testpiece because the
film is always on one side of the wall and the source on the other side. Figures 56(h), 56(j), and 56(k) illustrate alternative
arrangements for single-wall, single-image radiography. Shims, when used, need only compensate for any weld
reinforcement. When the radiation source is external, as in Fig. 56(h), location markers should be placed on the outside
surface for assurance that the correct overlap between adjacent exposures has been achieved. There should be a minimum
of one penetrameter and one set of view and location markers per film, except that there should be three or more
penetrameters and sets of markers (spaced equally around the circumference of the pipe) when a 360° simultaneous
exposure is made on a single strip of film, as shown in Fig. 56(k). A minimum of three penetrameters is needed for
assurance that the radiation source was actually located on the central axis of the cylinder and that equal intensity of
radiation was incident on the entire circumference. When a 360° simultaneous exposure is made on overlapping pieces of
film, not only should penetrameters be placed so that one appears on each piece of film, but also view markers and
location markers should be placed so that they coincide with the regions of overlap between adjacent pieces of film.
Radiography of Flanges. Although single-image techniques (especially the single-wall, single-image technique) are
ordinarily used with a normal (vertical) viewing direction, there are applications in which an offset view is advantageous.
Three setups for the single-image radiography of flanged pipe using offset views are illustrated in Fig. 56(m), 56(n), and
56(p). The principles of location-marker and penetrameter placement are similar to those previously discussed for normal
views; the only difference is that extra precautions must be taken to ensure that the projected images of markers or
penetrameters do not fall on the image of any region being inspected.
Radiographic Inspection
Revised by the ASM Committee on Radiographic Inspection
*
Control of Scattered Radiation
Although secondary radiation can never be completely eliminated, numerous means are available to reduce its effect. The
various methods, which are discussed below in terms of x-rays, include:
• Use of lead screens
• Protection against back scatter and scatter from external objects
• Use of masks, diaphragms, collimators, and filtration
Most of the same principles for reducing the effect of secondary x-rays apply also to -ray radiography. However,
differences in application arise because of the highly penetrating characteristics of gamma radiation. For example, a mask
for use with 200-kV x-rays could be light enough in weight for convenient handling, yet a mask for use with cobalt-60
radiation would be much thicker, heavier, and more cumbersome. In any event, with either x-rays or -rays, the means
for reducing the effects of secondary radiation must be selected with consideration of cost and convenience as well as the
effectiveness.
Lead screens placed in contact with the front and back emulsions of the film diminish the effect of scattered radiation
from all sources by absorbing the long-wavelength rays. They are the least expensive, most convenient, and most
universally applicable means of combating the effects of secondary radiation. Lead screens lessen the amount of
secondary radiation reaching the film or detector, regardless of whether the screens increase or decrease the intensity of
detected radiation. (The intensifying effect of lead screens is discussed in the section "Image Conversion Media" in this
article.)
Sometimes, the use of lead screens requires increased exposure time (or image processing in the case of real-time
monitoring). If high radiographic quality is desired, lead screens should not be abandoned merely because the photon
energy is so low that they exhibit no intensifying action. However, at a sufficiently low photon energy, depending on the