Fig. 56
Correct placement of view markers, location markers, and penetrameters for radiographic inspection.
Dimensions given in inches
Even though the following discussion and Fig. 56 illustrate the placement of markers and penetrameters on weldments,
similar locations for markers and penetrameters can be used on testpieces that do not contain welds. In all arrangements,
penetrameters should be placed in the outer cone of the radiation beam.
Radiography of Plates. Figure 56(a) illustrates three alternative arrangements of penetrameters and identification
markers for the radiography of a weld joining one plate to another plate of different thickness. In all three arrangements,
the identification markers and penetrameters are placed parallel to the weld. View markers and penetrameters are usually
placed 3 to 20 mm ( to in.) from the edge of the weld zone, but no more than 40 mm (1 in.). Testpiece
identification markers, however, can be placed farther away if necessary to ensure that their image is outside the image of
the weld zone in the processed radiograph. Identification markers are usually placed on the film but view markers should
be placed on the surface of the testpiece closest to the radiation source so that correct overlap between adjacent exposures
can be verified. (If the view markers were located on the film side, a portion of the testpiece directly above the view
markers could be missed even though the images of the markers appeared in adjacent radiographs.)
In Fig. 56(a), the preferred setup (setup 1) has two penetrameters located on the thinner plate. In the alternative setups
(setups 2 and 3), two penetrameters are located on the thicker plate (setup 2) or one penetrameter on each plate (setup 3).
Shims made of an alloy that has the same absorption characteristics as the weld metal are used under the penetrameters in
each instance to compensate for any difference between the thickness of the weld zone, including reinforcement, and the
thickness of the plate on which the penetrameter is located. Any shim used should be larger than the penetrameter placed
on it, so that the image of the penetrameter can be clearly seen within the umbral image of the shim. Also, the direction of
radiation with respect to shim and penetrameter location should be considered, especially with thick shims or
penetrameter blocks, to ensure that the shim properly represents the effective penetrated thickness of the testpiece.
Some codes and specifications require that the image of a penetrameter be used to evaluate the quality of only that portion
of the radiographic image of the testpiece that has similar photographic density. Strict limits can be placed on the
allowable density difference between penetrameter image and testpiece image. For this reason, it may be necessary to use
two or more penetrameters to evaluate image quality in different regions on the radiograph. When plaque-type
penetrameters are used, plaques of different thickness are used for different regions, depending on testpiece thickness in
each region.
Radiography of Cylinders. Figures 56(b), 56(c), and 56(d) illustrate alternative locations for markers and
penetrameters for the double-wall, double-image radiography of hollow cylinders or welded pipe. These alternatives can
be used for either normal or offset (corona) views. When the penetrameter is placed on the cylinder itself, as shown in
Fig. 56(b), or on a short section of pipe having the same diameter and wall thickness as the pipe being inspected, as
shown in Fig. 56(c), any shim that is used under the penetrameter should be only thick enough to compensate for weld
reinforcement; that is, twice the nominal reinforcement for a normal view, but equal to the nominal reinforcement for an
offset (corona) view. If the testpiece is a plain cylinder or if a circumferential butt weld is flush with the surface, no shim
is needed. When a penetrameter block is used to provide equivalent penetrated thickness under the penetrameter, the
block thickness should equal twice the nominal wall thickness of the cylinder plus twice the nominal weld reinforcement.
Also, the penetrameter block should be set on a block of Styrofoam or similar nonabsorbing material so that the upper
surface of the penetrameter block is aligned with the upper surface of the pipe, as shown in Fig. 56(d). When a short
section of pipe is used under the penetrameter (Fig. 56c), the radiation source should be centered between the pipe being
inspected and the short pipe section; otherwise, the radiation source should be centered above the pipe being inspected.
To ensure that the penetrameter image is within the umbral region of the image of a shim or penetrameter block, the
penetrameter should be aligned with the edge of the shim or block closest to the central beam of radiation.
Figures 56(e), 56(f), and 56(g) illustrate alternative setups for the double-wall, single-image radiography of hollow
cylinders or welded pipe. These alternatives are suitable for both normal and offset (corona) views. As for a double-wall,
double-image technique, the penetrameter can be placed on the pipe itself--on a short section of pipe being inspected or
on a penetrameter block. The setup illustrated in Fig. 56(e) can be used when there is access to the inside of the pipe for
placement of the penetrameter. When a shim is used under the penetrameter (Fig. 56e and f), it should be equal to the
height of nominal weld reinforcement, regardless of the view that is used. When there is no reinforcement, no shim is
needed. If the penetrameter is placed on a penetrameter block, as in Fig. 56(g), the block should be equal to twice the