Affinity Coelectrophoresis 409
agitating the pipet tip back and forth while rapidly drawing the solution in and out of the
tip 10 times, taking care not to generate bubbles in the mixture. Then add the mixture
(now totaling 250 µL) to lane 9, which is the rightmost of the nine lanes of the gel, taking
care to overfill the lane slightly. Any excess agarose that spills over into the zone between
sample lanes will quickly solidify and will not interfere with the running of the gel,
whereas underfilling the protein lanes results in anomalous GAG/PG migration through
the gel. Move on to the remaining samples, filling the lanes in the following order: 7, 5, 3,
1, 8, 6, 4, and 2. The key to success during this step is to mix the samples thoroughly but
to make sure to work quickly enough that the agarose–protein mixture does not gel in the
tube. If need be, the mixing can be done while the sample tube remains partially sub-
merged in the water bath.
9. Prepare the PG or GAG sample by mixing the labeled material in sufficient quantity and
activity that it can be detected (usually at least 10,000–20,000 cpm of
35
S-radiolabeled
material is sufficient for a good 3-d exposure in a phosphorimager cassette); tracking
dye(s) (we use 0.05% each of bromophenol blue and xylene cyanol); sucrose so that the
sample will sink through the RB during gel loading (5% w/v), and enough RB to bring the
volume to approximately 100 µL/gel (see Note 3). Once the labeled sample is mixed, it
should be vortexed and any insolubles pelleted at 13,000g for 2 min. When adding the
sample to the gel, be careful not to disturb any pellet that may be at the bottom of the tube.
10. The electrophoresis apparatus is next prepared. Make sure the apparatus is resting prop-
erly on a level stir plate. Attach the cooling hoses to a water source, either cold tap water
or a circulating water chiller set to 10°C, and turn on the water. Fill the apparatus with
cold running buffer to roughly 3 mm above the platform, then place the ACE gels with the
PG/GAG lane end of the gel oriented closest to the cathode, and use glass microscope
slides to hold them in place. Turn on the stir plate to purge air trapped in the buffer
recirculator, then turn the stir plate off. Add more buffer if needed, to be sure that the gels
are covered by about 3 mm of buffer. Using a pipetter with a 200-µL gel-loading pipet tip,
load the labeled sample into the lane, drawing it across the gel while discharging the
sample, so that it fills evenly. Take care not to nick the gel with the pipet tip during
sample loading—this leads to an uneven sample front and/or rapid loss of the sample into
the running buffer during electrophoresis (see Note 4).
11. Attach the cover to the gel box, plug the leads into the power supply, and turn it on to
deliver 60–80 V. After the tracking dye has entered the gel (in approximately 5 min), turn
on the stir plate to an intermediate setting but take care that the resulting agitation of the
buffer does not cause the gels to float away from the platform.
12. Run the gel for the appropriate time. For example, for gels containing protein-containing
lanes 15-mm-long, heparin samples should generally run for about 1.0 h, by this time the
heparin should have migrated most if not all of the way through the lane. PGs and other
GAGs may migrate at different rates depending on their relative sizes, charges, etc. The
position of the tracking dyes can be used as a guide to assess when the gels are finished—
for example, the dyes that we use (see step 9) typically move at approximately one-third
the rate of heparin in the electrophoretic field. After electrophoresis, turn off the power
supply and remove the gel box cover. Remove the gels and place them on an elevated
surface that will not impede the flow of warm air around them. We use 8-cm-high plastic
test tube racks placed about 30 cm in front of a warm air source supplied by a small
personal space heater, such as the Holmes (HFH 195, 1500 W). Allow the gels to dry at
the high heat setting for at least 8.0 h; they are dry when the agarose has flattened to a thin
clear sheet that is not sticky to the touch.