
896 Charged Particle and Photon Interactions with Matter
and covers the pseudo-rapidity range of 3.2 < │η│ < 4.9. It consists of three consecutive modules
along the beam line. The modules consist of a tungsten matrix in FCal2 and FCal3 (copper matrix
in FCal1) housing cylindrical electrodes, consisting of copper rods. The LAr gaps are very small:
0.250, 0.375, and 0.500mm, for FCal1, FCal2, and FCal3, respectively. The readout granularity
is Δη × Δϕ = 0.2 × 0.2. The design of the FCal was constrained by the high radiation level in the
most forward region. The detailed studies of the construction and performance are described in
Aubert etal. (2005).
31.3.4 lxe Scintillation caloriMeter (Meg experiMent)
The MEG experiment searches for a rare muon decay, μ
+
→ e
+
γ, which is forbidden in the standard
model. Several new theories beyond the standard model predict the branching ratio of the decay just
below the current experimental limit, that is, 1.2 × 10
−12
(Brooks etal., 1999). The MEG experi-
ment aims a sensitivity of 10
−14
(Mori, 1999), where most predictions are covered. The discovery of
the muon decay is a probe for new physics beyond the standard model. A μ
+
→ e
+
γ decay event is
characterized by the clear two-body nal state where the decay positron and the γ ray are emitted in
opposite directions with energies equal to half the muon mass (E = 52.8MeV). While positrons of
this energy are abundant from the standard Michel decay of muons, γ rays with such high energies
are very rare. Therefore, the key requirement for the MEG detector is a high energy resolution
for γ rays, since the accidental background rate decreases, at least, with the square of the energy
resolution
(Mori, 1999).
The
excellent properties of LXe as a scintillator and the developments of new PMTs with high
quantum efciency, fast timing, and their possible operation in LXe motivated the choice of the
detector for the MEG experiment. With the Xe scintillation light detected by a large number of
PMTs immersed in the liquid volume, the detector’s energy resolution is proportional to
N
pe
,
where N
pe
is the number of photoelectrons, and was estimated to be better than 1% (1σ) for 52.8MeV
γ rays (Doke and Masuda, 1999). In addition, the γ ray interaction points can also be determined
from the light distribution on each PMT (Mori, 1999). Following the R&D with a small proto-
type (Mihara etal., 2002) and with further improvement of the metal channel-type PMT, a large
LXe prototype detector with 228 PMTs of 2in. was constructed and tested with 10–83MeV γ rays
(Mihara etal., 2004). The detector and associated cooling and purication apparatus, shown sche-
matically in Figure 31.15, were tested at the Paul Scherrer Institute in Switzerland where the nal
MEG experiment is located. A pulse tube refrigerator (PTR) was used for the rst time to condense
Xe gas and maintain the liquid temperature stable over a long period. The PTR, optimized at the
LXe temperature, had a cooling power of 189W (at 165K) (Haruyama, 2002). The prototype was
large enough to test both the reliability and stability of the cryogenic system based on the PTR, and
the
efciency of the purication system required for the calorimeter’s performance.
While
the transmission of the scintillation light is affected by Rayleigh scattering, this process
does not cause any loss of light, and hence does not affect the energy resolution. Two types of cir-
culation systems for Xe purication were developed to improve the light attenuation length. One is
gaseous purication using a heated metal getter, in which the evaporated Xe is brought to the hot
getter system, and then the Xe is recondensed in LXe in the detector (Baldini etal., 2005a). This
method is effective in removing all types of impurities, but is not efcient for water, which was iden-
tied as the major contributor to light absorption, because its vapor pressure at the LXe temperature
is too low. Furthermore, the purication speed is limited by the cooling power. The other method
was developed specically to remove water at a much higher rate (100L/h) by circulating LXe with
a cryogenic centrifugal pump. Water is efciently removed by a lter containing molecular sieves.
With this method, the total impurity concentration was reduced in 5h from 250 to 40 ppb in the total
100 L
LXe volume of the MEG prototype detector (Mihara et al., 2006).
A
novel calibration technique was developed and applied to the prototype detector. A lattice of
α point sources deposited on thin (100 μm diameter) gold-plated tungsten wires was permanently