PART IV Pulse Doppler Radar
256
Phasor Representation of the Samples. Everything we
have seen by viewing the detector output at a point on a
range trace corresponding to a particular target’s range can
be presented much more conveniently in a phasor diagram,
such as that shown in Fig. 8. Moreover, a phasor diagram
presents the outputs of both the I and Q detectors simulta-
neously. The length of the phasor corresponds to the ampli-
tude (A) of the target’s echoes. The angle (φ) which the pha-
sor makes with the x axis, corresponds to the radio-fre-
quency phase of the echoes relative to the reference signal.
The length, x, of the projection of the phasor on the x axis
corresponds to the output of the in-phase (I) detector; the
length, y, of the projection on the y axis corresponds to the
output of the quadrature (Q) detector.
The phasor rotates at the target’s apparent doppler fre-
quency, i.e., its true doppler frequency or true doppler fre-
quency plus or minus an integer multiple of the sampling
rate. If this frequency is positive (greater than the reference
frequency), rotation is counterclockwise (Fig. 9); if nega-
tive, rotation is clockwise. The amount that the phasor
steps ahead from sample to sample (∆φ) is 2π radians
(360°) times the doppler frequency times the length of the
sampling interval:
∆φ = 2π f
d
T
s
where f
d
is the apparent doppler frequency and T
s
is the
sampling interval. If the sampling rate is the PRF (as it gen-
erally is in all-digital signal processors),
2
T
s
is the interpulse
period, T.
What the Filter Does
Digital filtering is simply a clever way of adding up (inte-
grating) successive samples of a continuous wave so that
they produce an appreciable sum only if the wave’s frequen-
cy lies within a given narrow band—i.e., produce a sum
equivalent to the output which would be produced if the
continuous wave were applied to a narrow band analog fil-
ter. If the variation in amplitude from sample to sample cor-
responds closely to the resonant frequency of the equivalent
analog filter, the sum builds up; otherwise, it does not.
What the filter does, in effect, is project the x and y com-
ponents of the phasor representation of the samples onto a
rotating coordinate system (i and j in Fig. 10). The rate at
which the coordinates rotate—number of revolutions per
second—is made equal to the center frequency of the band
the filter is intended to pass. This rate, f
f
, can be thought of
as the filter’s resonant frequency: the frequency to which it
is “tuned.”
8. If the sine wave is represented by a phasor (A), the I compo-
nent is the projection of the phasor on the x axis and the Q
component is the projection of the phasor on the y axis.
9. The amount that the phasor steps ahead from sample to sam-
ple (∆φ) is proportional to the target’s doppler frequency.
10. The filter projects the x and y components of the phasor, A
n
,
onto a coordinate system (i, j) that rotates at the frequency to
which the filter is “tuned.”
2. In certain applications digital
doppler filtering may be pre-
ceded by some analog filter-
ing (for clutter reduction),
which converts the return to a
CW signal. The sampling rate
then is generally not the PRF.
Click for high-quality image