For metals, excluding shapes such as tubes, the movement of the dial gage caused by the elasticity of the metal
being tested is small and not considered to be a problem. Elasticity may reach considerable proportions with
plastics. In addition to the spring of the tester frame, elasticity may prevent full application of the major load
because of limitations in the design of the tester.
The limitation of the standard model Rockwell tester is considered to be 150 dial gage divisions under a 150
kgf load. This figure represents the number of divisions of travel on the dial gage, when the major load is
applied, due to penetration into the material under tests, spring of the frame, penetrator, plunger rod system, and
elasticity of the material under test. Special Rockwell testers, designated as “PL” models, increase this
limitation to 250 divisions under a load of 150 kgf.
To determine whether the machine limitation is being exceeded and the major load is being fully applied, the
major load can be tested in the following manner. With the major load still applied, an additional load can be
applied by manually exerting pressure on the weights on the machine; the dial gage needle then should indicate
additional penetration. If not, the full major load might not be acting (due to reaching limit of depth of
indentation), and faulty readings can result. In this instance, the manufacturer should be contacted.
Use of the Alpha Scale. A variation of the standard Rockwell test is often used for testing plastics. It is referred
to as the alpha Rockwell hardness number in Procedure B of ASTM D 785 (Ref 2). The advantage of the alpha
scale is that it covers the range of plastics.
The standard Rockwell tester is used with a major load of 60 kgf and 12.7 mm (½ in.) ball penetrator. The test
is made by applying the minor load in the usual manner, setting the dial to “set,” and applying the 60 kgf major
load for 15 s. With the major load applied, the number of divisions the penetrator has traveled from “set” is read
on the dial gage. From this reading, the spring of the tester is subtracted, determined under the major load of 60
kgf, and the remainder is subtracted from 150.
The spring of the machine, known as the “spring constant,” is determined as follows:
• Place a soft copper block of sufficient thickness and with plane parallel surfaces on the anvil in the
normal testing position.
• Raise the sample and the anvil by the capstan screw until the large pointer is at the set position.
• Apply the major load by tripping the load release lever.
The dial gage then will indicate the vertical distance of indentation, the spring of the machine frame, and any
other elastic compressible deformation of the plunger rod system and penetrator. This operation should be
repeated several times without moving the block. However, the dial must be reset after each test while under
minor load until the deflection of the dial gage becomes constant—that is, until no further indentation takes
place, and only the spring of the instrument remains. This value, in terms of dial divisions, is the spring
constant.
Durometer Testing. The durometer is a well-known and widely used instrument for measuring hardness of
virtually all types of plastics, rubbers, and various rubberlike materials. The durometer measures hardness by
means of an indentation much like that used in hardness testing of metals. The indenters used in durometers,
however, are spring loaded rather than forced by weights. Nonmetallic materials, similar to metals, vary greatly
in hardness, thus requiring a variety of test instruments. Several types of durometers accommodate the full
range of hardness, and special instruments are available for testing O-rings and extremely thin materials. The
various types available are listed in the left column of Table 14; however, only two (A and D) are covered in
ASTM D 2240 (Ref 8).