
The analysis of hydrocarbon gases
31
There have been many references to columns and column combinations for
separation of all C, saturates and unsaturates
[17-191.
One of the earliest [20]
referred to the use of activated alumina, the surface of which was strongly ad-
sorptive to unsaturated hydrocarbons. When fully activated, retention times for
unsaturated hydrocarbons were excessive, and their peak shapes were poor,
so
a
degree of deactivation was recommended, using water or a combination of water
and silicone oil. This gave good separation, but was only consistent over a few
hours, as the water was stripped off by the dry carrier gas and the relative reten-
tions
of
saturated and unsaturated hydrocarbons changed significantly. It was
usually necessary to precede a day’s work with the injection of a relatively large
amount of liquid water, which would distribute itself over the alumina and allow
several hours of reasonably consistent separations.
Capillary columns packed with alumina were demonstrated [21] shortly after
the first reference, and later applied to refinery gas analysis [22]. In this applica-
tion, a humidification device was used for the carrier gas. This consisted of a
length of tubing packed with copper sulphate crystals, which, at ambient tem-
perature, assured a constant water content in the carrier gas, and hence a constant
polarity. In fact, with this device, there is the mechanism for fine-tuning the col-
umn behaviour. The vapour pressure of water above copper sulphate crystals is
constant at a constant temperature, and
so
the partial pressure of water in carrier
gas is controlled by adjustment of the total pressure in the humidifier. With the
arrival of porous layer open tubular (PLOT) capillary columns, alumina was
used in this form. The separations which had been incomplete with packed col-
umns were now baseline to baseline.
These columns were rapidly seen to give superior performance to any of the
alternatives, and became the state-of-the-art solution to the problem. The need to
maintain a constant level of moisture in the carrier gas was inconvenient, and
so
alternative methods for deactivation were studied. The use of inorganic salts
proved to be effective, and potassium chloride treatment was the first such to be
commercially successful. KCI deactivated alumina, used with dry carrier gas,
gives excellent separation, but still has an affinity for moisture, which can alter
both absolute and relative retentions. When not in use, therefore, it is recom-
mended that the column should be kept
at
a relatively high temperature (e.g.
150OC) to avoid adventitious moisture. Columns are available with 0.32 mm in-
ternal diameter, with a 5pm layer, or with 0.53 mm internal diameter, with a
10pm layer.
Figure 1.22 is a chromatogram of hydrocarbons on KCI-deactivated alumina.
The separations are good, such that most components could be measured even if
they were trace impurities. The separations which could be improved are those
of acetylene and propadiene from the C, saturates, and of 1-butene from trans-2-
butene. Further studies of alternative deactivating agents [23] showed that the
use of sodium sulphate rather than potassium chloride gave a phase with rather
References
p.
40