backscattering. This may be mounted on the specimen holder or, more conven-
iently, on a retractable arm (a standard feature of EMPs, but not SEMs).
As noted in Section 3.4, there is a direct relationship between the strength of
the condenser lenses and the beam current. The required current may therefore
be obtained by adjustment of the condenser lenses, which, in a computer-
controlled instrument with current monitoring, can be done automatically by
means of a software command. (This also has the effect of changing the final
beam diameter.)
Drift in beam current as a function of time is caused mainly by movement of
the tip of the filament, which can be corrected using beam alignment coils (as
discussed in the previous section). Such drift is a potential source of error
in automated quantitative X-ray analysis sessions, which may extend over
several hours. One solution to this problem is to insert the Faraday cup before
each measurement and normalise the intensities. A better alternative, however,
is to use a regulating system that continuously adjusts the condenser lenses to
maintain a constant current. This requires a double aperture, in which part of
the beam passing through the first aperture is intercepted by the second
(smaller) one, and the current collected by this is used as input to a feedback
amplifier. (It is assumed that this current is proportional to that passing
through the second aperture.) This is a normal feature of EMPs but not SEMs.
3.7 Beam scanning
Scanning images are produced by sweeping the beam across the sample in a
television-like ‘raster’ while displaying the output of an electron detector on
the screen of a synchronously scanned visual display unit (VDU), as shown in
Fig. 3.9. The electron beam is deflected by means of coils located above the
final lens, which enable the beam to be ‘pivoted’ about the centre of the lens.
These are supplied with a ‘sawtooth’ current waveform derived from line and
frame scan generators. The ratio of frame and line scan frequencies determines
the number of lines in the image (typically 500–2000).
Instead of scanning a rectangular raster, the beam can be swept along a
single line by using only one set of coils, in order to produce a line plot, which is
useful for some purposes. The deflection system can also be used to move the
beam around in ‘spot’ mode for X-ray or other forms of analysis on selected
points.
‘Analogue’ scanning systems have been superseded by digital systems in
which the beam deflection is computer-controlled via a digital-to-analogue
converter (DAC) and the output from the signal detector is converted into
digital form by an analogue-to-digital converter (ADC), so that the intensity at
3.7 Beam scanning 29