CHAPTER 4. INSTRUMENTATION
132
detectors, the high density of BGO giving it a very effective stopping power for
511 keV photons. Sodium iodide (NaI), the detector of choice for gamma
cameras and SPECT systems, is also used; however, it has several properties
which make it less attractive than BGO. It does have a higher light output and
consequently better energy resolution than BGO (~10% compared with ~20%)
but, besides its significantly lower stopping power, it is hygroscopic.
Dedicated PET systems may use lead collimation in the form of 1 mm x 65
mm lead septa, which significantly reduces scattering and random events.
Recent developments have involved removal of these septa, so that there is no
collimation within the imaging volume of the scanner, necessitating a 3-D recon
-
struction of the data. This can potentially improve sensitivity by a factor of six.
The differences between PET and SPECT in performance and applica-
tions are decreasing. Dedicated multidetector SPECT instruments with purpose
designed collimators, for example fanbeam collimators, now permit SPECT
studies to be performed with improved sensitivity compared with that of single
head SPECT and comparable resolution to that of PET, although sensitivity is
still lower than that of PET.
4.4.3. Hybrid PET–SPECT instrumentation
Although gamma cameras were used in the early days of PET, manufac-
turers have recently introduced commercial gamma cameras based on
coincidence detection systems. Dual head gamma cameras with opposing heads,
originally designed for multidetector SPECT, can be used with additional
coincidence circuitry to detect positron events in exactly the same way as PET
detectors. The clear advantage of these systems is their relatively low cost and
the fact that the instrument can be used for either SPECT or coincidence
detection (CD).
The term CD is often used to differentiate these systems from dedicated
PET systems, although a more appropriate term is ‘hybrid PET–SPECT
systems’. The only real difference is that the dual detectors must be rotated
around the patient, as in normal SPECT acquisition, whereas dedicated PET
systems are designed with a complete (or partial) ring of detectors that surround
the patient.
The main differences in the design of a hybrid PET–SPECT system,
compared with a dual head SPECT system, are as follows.
Coincidence circuitry must be added so that the two opposing detectors
can detect the two annihilation photons in coincidence, i.e. within a very short
time of each other. It is this coincidence that defines the path along which the
photons must have travelled, eliminating the need for a conventional collimator.