A. Sugimura and G. R. Luckhurst 202
field-induced alignment, have been well-characterized.
14,31,35
To obtain
the simplest deuterium NMR spectrum the sample was deuteriated in the
α
-position of the pentyl chain.
13
As we shall see the deuterium NMR
spectrum of a monodomain of 5CB-d
2
contains a single quadrupolar
doublet because the dipolar coupling between the two deuterons is
negligible in comparison with the linewidth. In principle, this could be
reduced by decoupling the dipolar interactions to neighbouring protons
which broaden the lines inhomogeneously.
36
The sample of 5CB-d
2
was
contained in a cell 56.4
µ
m thick composed of two glass slides each
coated with indium oxide to form the electrodes, although some
measurements were also made with a cell
97
µ
m thick. The surfaces of
the electrodes were not treated in any way and so the surface anchoring
strength was of the order of 10
-7
Jm
-2
, corresponding to weak anchoring.
The sinusoidal electric field was produced using a function generator,
Wave Factory WF1943, NF Electronic Instruments and a high power
amplifier, model 4005 NF Electronic Instruments.
The deuterium NMR spectra were measured using a JEOL Lambda
300 spectrometer which has a magnetic flux density of 7.05T. A
quadrupolar echo sequence, sketched in Fig. 4, was used to record the
free induction decay (FID) which was then Fourier transformed to obtain
the conventional frequency domain spectrum; also shown in this figure is
the sinusoidal electric field. This field was applied for a total time, t
a
,
corresponding to an integer number of cycles and finishing after the FID
acquisition was complete, the dead time of the receiver coil,
t
D
, was
about 10
µ
s, the time between pulses was 38
µ
s, each pulse was 7.5
µ
s
long, the acquisition time for the FID was 2.56 ms and finally the delay
time, PD, was given values in the range 50 to 200 ms with the longer
times being used at the lower frequencies. This delay allows the director
to be realigned parallel to the magnetic field following the removal of the
electric field and, incidentally, allows the deuterium spins to relax back
to their equilibrium distribution. The experiment was performed by
activating a trigger to switch on the electric field, at the same time the
trigger initiates the sequence which leads to the measurement of a single
FID. The time at which the measurement takes place, following the
application of the field, is controlled by the interval PI3 (see Fig. 4)
although the total time to the acquisition of the FID is the sum of PI3,
t
D
,