Practical Issues
203
gasifiers, the tip of the assembly is typically located slightly withdrawn into the
wall, so as to protect it from slag or other erosion damage. The actual temperature
measured is closer to that of the refractory rather than that of the reactor core and
thus highly dependent on the extent of the depth of withdrawal from the reactor space.
Despite these disadvantages, platinum-rhodium thermocouples are still the most
common device currently used for gasifier temperature measurement. It is accepted
that real accuracy of temperature measurement is less important than consistency.
For the gasifier operator, who has set his feed inputs on the basis of other parameters,
a continuous and steady temperature reading is more important than the absolute
value shown.
Nonetheless, investigations continue to develop alternative methods, not least
because in oil gasification thermocouple life can be run-length determining. Systems
under consideration or development include the following:
•
Pyrometers
. Texaco has used a pyrometer in its pilot unit for several years, and
this will shortly be tested in a commercial environment (Leininger 2002). The
principle advantage of such a system is that the sensor is located outside the reactor
and not subject to the harsh environment. The necessity to ensure pressure integrity,
including a high pressure nitrogen purge, does make it expensive, however. Interest-
ingly, the actual temperature measured is dependant on the gasifier fuel. With gas
firing, the visible path reaches to the opposite wall of the reactor, so that the
temperature measured is that of the refractory. Depending on the degree of solids
in the reactor, the visible path may reach only to the center, that is, the hottest location
in the reactor, or even less where the temperature is cooler again. Furthermore, the
nitrogen purge can cool the slag around the line of sight of the pyrometer leading to
a loss of reading. Interruption of the nitrogen purge can solve this problem online,
a distinct advantage over thermocouples, which generally require a reactor shutdown
for replacement. Interpretation of such a loss of reading does require additional
temperature measurements, so that any commercialization of pyrometry is likely
to be in addition to rather than as a replacement of thermocouples.
•
Steam make in the membrane wall
. This measurement is of course limited to reac-
tors having a membrane wall or water jacket. As was already mentioned in Section
6.4.2, the steam make in the membrane wall is a valuable indicator for both the heat
loss through the wall and for the reactor temperature. It has the advantage of being
an integral measurement of the temperature anywhere near the wall of the reactor.
It is fast, with a response time of less than a minute, and very reliable. It does not,
however, provide local temperature values.
•
Heat flux measurement
. This measurement comprises installing a small piece of
membrane wall in the wall of a reactor and measuring the increase in water
temperature of a known flow of water through the membrane wall. It can give a
fast—10–30 second response time—indication of the local temperature.
•
Microwave-based measurements
have also been considered.
•
Other devices
. A new system based on temperature-dependant changes in the optical
properties of single-crystal sapphire is under development. Despite some promising