32 2 Energy-Loss Instrumentation
distance v
x
from the exit of the magnet; see Fig. 2.2a. This focusing action occurs
because electrons with positive γ
x
travel a longer distance within the magnetic field
and therefore undergo a larger angular deflection, so they return toward the optic
axis. Conversely, electrons with negative γ
x
travel a shorter distance in the field are
deflected less and converge toward the same point I
x
. To a first approximation, the
difference in path length is proportional to γ
x
, giving first order focusing in the x–z
plane. If the edges of the magnet are perpendicular to the entrance and exit beam
(ε
1
= ε
2
= 0), points O, I
x
, and C (the center of curvature) lie in a straight line
(Barber’s rule); the prism is then properly referred to as a sector magnet and focuses
only in the x-direction. If the entrance and exit faces are tilted through positive angles
ε
1
and ε
2
(in the direction shown in Fig. 2.2a), the differences in path length are less
and the focusing power in the x–z plane is reduced.
Focusing can also take place in the y–z plane (i.e., in the axial direction, paral-
lel to the magnetic field axis), but this requires a component of magnetic field in
the radial (x) direction. Unless a gradient field design is used (Crewe and Scaduto,
1982), such a component is absent within the interior of the magnet, but in the fring-
ing field at the polepiece edges there is a component of field B
n
(for y = 0) that is
normal to each polepiece edge (see Fig. 2.2a). Provided the edges are not perpen-
dicular to the optic axis (ε
1
= 0 = ε
2
), B
n
itself has a radial component B
x
in
the x-direction, in addition to its component B
z
along the optic axis. If ε
1
and ε
2
are positive (so that the wedge angle ω is less than the bend angle φ), B
x
> 0for
y > 0 and the magnetic forces at both the entrance and exit edges are in the negative
y-direction, returning the electron toward a point I
y
on the optic axis. Each boundary
of the magnet therefore behaves like a convex lens for electrons traveling in the y–z
plane (Fig. 2.2b).
In general, the focusing powers in the x- and y-directions are unequal, so that
line foci I
x
and I
y
are formed at different distances v
x
and v
y
from the exit face; in
other words, the device exhibits axial astigmatism. For a particular combination of
ε
1
and ε
2
, however, the focusing powers can be made equal and the spectrometer is
said to be double focusing. In the absence of aberrations, electrons originating from
O would all pass through a single point I, a distance v
x
= v
y
= v from the exit. A
double-focusing spectrometer therefore behaves like a normal lens; if an extended
object were placed in the x–y plane at point O, its image would be produced in the
x–y plane passing through I. But unlike the case of an axially symmetric lens, this
two-dimensional imaging occurs only for a single value of the object distance u.Ifu
is changed, a different combination of ε
1
and ε
2
is required to give double focusing.
Like any optical element, the magnetic prism suffers from aberrations. For
example, aperture aberrations cause an axial point object to be broadened into an
aberration figure (Castaing et al., 1967). For the straight-edged prism shown in
Fig. 2.2a, these aberrations are predominantly second order; in other words, the
dimensions of the aberration figure depend on γ
2
x
and γ
2
y
. Fortunately, it is possible
to correct these aberrations by means of sextupole elements or by curving the edges
of the magnet, as discussed in Section 2.2.
For energy analysis in the electron microscope, a single magnetic prism is the
most frequently used type of spectrometer. This popularity arises largely from