Chromatography 95
4.3.4 An ion
-
exchange chromatography run
The choice of the type of exchanger, cation
-
or anion
-
exchanger,
may be arbitrary, in the absence of any knowledge of the characteristics
of the protein of interest, e.g. its pI and/or its pH stability range.
As
a
first approach, it is generally best to choose a pH, within the protein’s
stability range, where it will adsorb to the ion
-
exchanger. Usually, this
means that an anion
-
exchanger should be used at pH values above the pI
of the protein and a cation
-
exchanger at a pI below the pI. (It must be
realised, however, that the pI refers to an overall property of the
protein, whereas binding to an ion
-
exchanger is a function of the charge
on one surface of the protein. It is therefore possible to have a protein
bind to an anion
-
exchanger at a pH below its pI or to a cation
-
exchanger
at a pH above its pI. It becomes a matter of experimentally exploring
the behaviour of each new unknown protein.)
With the resin chosen and the column packed, it is necessary to
equilibrate the column with several column volumes (“colvols”) of
starting buffer, a buffer of low ionic strength and of a pH which will
promote binding of the protein of interest to the resin.
The sample
protein mixture must contain a low salt concentration, achieved by
equilibrating it with the starting buffer; either by dialysis, ultrafiltration,
molecular exclusion chromatography or, following TPP,
by simply re
-
dissolving the precipitate in the starting buffer.
The sample solution is applied to the column and chased with at least
2 colvols of starting buffer in order to elute the unbound fraction. The
A
280
may be monitored
during this process and elution with starting
buffer stopped once the A
280
returns to the baseline.
At this point a
buffer gradient may be applied. As a first approach, a gradient of
increasing ionic strength is the best choice, and is applicable to both
cation
-
and anion
-
exchange.
The resolution of peaks is a function of the steepness of the eluting
gradient. A shallow gradient gives better resolution, but takes more time,
so a trade
-
off must be made. With an unknown system, the best first
approach is to use a steep gradient, as this gives a quick assessment of the
number of peaks to be expected, and the separation can subsequently be
optimised. A suitably steep gradient for a first approach is M NaCl
in starting buffer, in 3 colvols. For subsequent optimisation the gradient
limits or the number of colvols can be altered, to change the gradient
slope.
It must be appreciated that the gradient is applied to the inlet side of
the column, whereas monitoring of the effluent is done on the outlet side.
There is thus a colvol difference between the influent and effluent