Temperature and Tracer-Gas Concentration. For leak testing, the temperature should be held constant, when
possible, preferably at the expected operating temperature of the item tested. Also, tracer-gas concentration should be as
high as possible to ensure sufficient detector response.
Vacuum Leak Testing Methods
The equipment most commonly used for the leak testing of vacuum systems involves the use of manometers, halogen
detectors, mass spectrometers, ionization gages, thermal-conductivity gages, residual gas analyzers, partial pressure
analyzers, and gas chromatographs. The methods used with this equipment are listed in Table 3. Methods that can be used
with both vacuum and pressure systems are discussed in detail in the section "Leak Testing of Pressure Systems Using
Specific-Gas Detectors" in this article.
Manometers. Testing for leaks in systems under vacuum can be done by observing the manometers installed on the
system. Although manometers are not as adaptable to locating a leak as regular leak detectors, some indication of leakage
can be attained. Because most vacuum gages will react or change their reading in the presence of a tracer gas, spraying
tracer gases over individual portions of the vacuum system will produce a gage response and will provide some indication
of leak location. The location can also be roughly established by isolating various portions of the system and observing
the gage response. The primary evidence of a leak in the system is failure of the system to achieve anticipated pressure
levels in a certain pumping time as based on past or calculated performance. However, consideration must be given to
other gas loads (virtual leaks) that may exist in the system. Sources of these gas loads are system contaminants, such as
water, outgassing of chamber surfaces and test items, and outgassing of entrapped areas such as unvented O-ring grooves.
Pressure-change gages having a pressure-change sensitivity of 0.13 mPa (10
-6
torr) are available for vacuum applications.
Ionization gages will measure vacuums to 1.3 pPa (10
-14
torr) and can be used to measure changes in pressure. The
sensitivity of all pressure-change methods is time dependent (that is, the mass changes per unit of time); therefore, the
longer the test duration, the more sensitive the method.
Halogen Detectors. Heated anode halogen detectors can also be used on vacuum systems. With special adapters, they
can be used down to 0.13 Pa (10
-3
torr). Leaks are detected by using an externally applied tracer gas.
Mass spectrometers are well suited for use as leak detectors in vacuum systems because the detector tube of a mass
spectrometer must be maintained at a vacuum. Also, mass spectrometers usually have built-in vacuum-pumping systems
that can be directly coupled to the system being tested. Helium mass spectrometers are most commonly used. Leaks are
detected by spraying or enveloping the outer surface of the system with helium.
Ionization Gages. In addition to detecting leaks in vacuum systems by pressure-change instruments, ionization gages
can be used to detect the presence of specific tracer gases. Because each gas traveling through the gage ionizes
differently, equal flows of different gases will produce different readings. Leak rates for molecular or laminar flow can be
related to ionization-gage readings in response to a tracer gas through:
(Eq 9)
(Eq 10)
where Q
a
is the flow of air (in torr liters per second), ΔR is the change in ionization-gage reading (in torr), S
a
is the
pumping speed (in liters per second), is the gage sensitivity factor, v = η
a
/η
t
(where η
a
is the viscosity of air and η
t
is
viscosity of tracer gas), and μ= (M
a
/M
t
) (where M
a
is the molecular weight of air and M
t
is the molecular weight of
tracer gas). Values of σ for ionization gages, and of the gas factors (σ - 1) and [(σv/μ) - 1], for several gases are as
follows: