The chemiluminescence detector
21
1
some impurities) in hexane solvent, with
10%
benzene. Figure 8.3B shows the
SCD chromatogram obtained under the proper gas flow rates (200 mumin
H2,
400
ml/min air). No response is observed for -130pg of hexane (2pl split
1O:l).
Under the chromatographic conditions employed, benzene co-elutes with thio-
phene, thus slightly distorting the thiophene peak on the SCD, however no
quenching is observed. Figure 8.3A shown the same test mixture using
200 ml/min H2 and 350 mumin of air. Under these more reducing conditions, a
negative response is observed for hexane and benzene and a greater response of
the detector for the sulfur compounds is obtained. Finally, Fig. 8.3C shows the
SCD response for the same test mixture using 200 ml/min
H2
and
500
ml/min of
air. The response for the sulfur compounds
is
greatly reduced and in some cases,
a positive response for the hydrocarbons can be observed under these more oxi-
dizing conditions. This example illustrates the importance of operating the detec-
tor at the proper gas flow rates.
When the SCD is operated at the proper gas flow rates, good day-to day re-
producibility can be obtained, however, contamination of the probe from column
bleed and septa bleed can cause a loss in sensitivity. Silicone compounds from
either GC columns or septa when high injection port temperatures are used can
deposit on the tip of the probe and decrease the response of the detector for sul-
fur compounds. In most cases, the contamination can be physically removed by
simply inserting a cleaning wire into the probe. To avoid the silicone bleed con-
tamination, the columns (and septa) should be well conditioned and the column
operated at as low of temperature as possible for a given analysis. For example,
most bonded methyl silicone columns can be operated at temperatures up to
275”C, without any bleed problem, but operation at higher temperatures such as
300°C can cause contamination of the probe and reduced sensitivity. It should be
noted that the SCD is more sensitive than the FID to silicone bleed and contami-
nation can occur even though no significant baseline rise is observed on the FID.
8.6
FLAMELESS SULFUR CHXMILUMINESCENCE
A
dedicated ceramic burner for the formation of sulfur monoxide has been re-
cently described by Shearer [25] and a schematic of the burner is shown in Fig.
8.4.
In this “flameless” SCD, hydrogen and air are mixed with the column efflu-
ent in a heated ceramic tube and all of the combustion gases are transferred to
the chemiluminescence detector. The typical gas flow rates for the burner are
100 mumin for hydrogen and
20
ml/min of air, which are outside the flammabil-
ity limits for H2/air. To initiate and sustain the combustion, the burner is heated
typically to 800-900°C and direct connection of the chemiluminescence detector
(and vacuum pump) to the burner results in a reduced pressure, typically
References
pp.
227-229