The scattering mechanism leads to a signi®cant build-up in the number of
detectable g rays, as shown in Figure 4.6(b) and discussed in Section 4.4.4.
There is consequently a net gain in the sensitivity of the counting system to
the tracer (Section 4.4.1). However, there is a price to pay. Spectral informa-
tion that can only be obtained from an analysis of countrates in the full
energy peak is largely lost. This is not important if, as is often the case, only a
single radiotracer is involved.
The measurement of scattered radiation involves integral counting techni-
ques. The lower level discriminator is adjusted to accept as much of the low-
energy radiation as possible while avoiding noise pulses which could become
signi®cant when detecting pulse heights below a few millivolts. The setting of
the upper level discriminator is adjusted to maximise the signal to back-
ground ratio. Readers are referred to Section 4.4.2 and Figure 4.7 for further
information.
System checks and calibrations should be undertaken under conditions as
close as possible to those in the ®eld. The location of the counting plateau,
and hence the operating voltage (Figure 8.2) depends on the characteristics of
the detector, and also on the energy of the absorbed radiation. It is therefore
common practice to establish the plateau position with an americium-241
source as its 59.5 keV g ray is comparable in energy to much of the scattered
radiation. Another option is to place a sample of the radiotracer in a thin lead
pot and set the plateau using the degraded radiation emitted from the surface.
Accurate ®eld measurements
Radioisotope investigations are becoming increasingly sophisticated, re-
quiring good quantitative data. Procedures for accurate activity measure-
ments in laboratories were discussed in Sections 6.3 and 6.4. The accuracy of
®eld measurements depends on the quality of the calibration of the detector
and this is normally limited by the need to reproduce in the laboratory the
counting geometry which is found in the ®eld.
Three classes of counting geometry may be distinguished:
. quasi-in®nite geometry, when the detector is completely immersed in a large
volume of liquid such as industrial tanks, rivers or lakes;
. quasi-planar geometry, when the detector is mounted, for instance, on a sled a few
centimetres above labelled sediment on the bed of an estuary; and
. quasi-linear geometry when the detector is mounted external to a pipeline.
These are limiting cases which are discussed in the following chapter and
illustrated in Figures 9.2(a) to 9.2(c) respectively.
8.2 Tracer applications in the ®eld 245