6 1 EPR (Electron Paramagnetic Resonance) Spectroscopy of Polycrystalline Oxide Systems
lower level, the odd electrons from the upper level give up the h ν quantum to
return to the lower level and satisfy the Maxwell – Boltzmann law. The release of
this energy occurs via a spin relaxation process, of which there are two types,
known as spin – lattice relaxation and spin – spin relaxation.
In the fi rst case, the energy is dissipated within the lattice as phonons, that is,
vibrational, rotational and translational energy. The mechanism by which this
dissipation occurs is known as spin – lattice relaxation. It is characterized by an
exponential decay of energy as a function of time. The exponential time constant
is denoted T
1e
and is called the spin – lattice relaxation time. In the second case the
initial equilibrium may also be reached by a different process. There could be an
energy exchange between the spins without transfer of energy to the lattice. This
phenomenon, known as spin – spin relaxation is characterized by a time constant
T
2e
called the spin – spin relaxation time.
When both spin – spin and spin – lattice relaxations contribute to the EPR line,
the resonance line width ( ∆ B ) can be written as
∆B
TT
∝+
11
12ee
(1.12)
In general, T
1e
> T
2e
and the line width depends mainly on spin – spin interac-
tions. T
2e
increases on decreasing the spin concentration, that is, the spin – spin
distance in the system. On the other hand when T
1e
becomes very short, below
roughly 10
− 7
sec, its effect on the lifetime of a species in a given energy level makes
an important contribution to the linewidth. In some cases the EPR lines are broad-
ened beyond detection.
T
1e
is inversely proportional to the absolute temperature ( T
1e
∝ T
− n
) with n
depending on the precise relaxation mechanism. In such a case, cooling the
sample increases T
1e
and usually leads to detectable lines. Thus quite often EPR
experiments are performed at liquid nitrogen (77 K) or liquid helium (4 K) tem-
peratures. On the other hand if the spin – lattice relaxation time is too long, elec-
trons do not have time to return to the ground state. The populations of the two
levels ( n
1
and n
2
) tend therefore to equalize and the intensity of the signal decreases,
being no longer proportional to the number of spins in the sample itself. This
effect, known as saturation, can be avoided by exposing the sample to low incident
microwave powers. This is an important consideration, particularly when estimat-
ing the number of spins in a paramagnetic system using a reference standard (see
Section 1.2.8 ).
1.2.3
The Nuclear Z eeman Interaction
If the interaction of the electron with an applied external magnetic fi eld were the
only effects detectable by EPR, then all spectra would consist of a single line and
would be of little interest to chemists. However, the most useful chemical infor-
mation that can be derived from an EPR spectrum usually results from nuclear