and X ray detectors beginning during the mid-1960s. High-resolution semi-
conductor g ray detectors are either crystals of lithium drifted germanium
(Ge(Li)) or high-purity germanium (HPGe) which make it possible to
routinely obtain spectra such as that of europium-152 (Figure 3.14(a)).
Europium-152 (T
1/2
= 13.51 y) undergoes a two-pronged decay (EC and
b
7
) followed by the emission of g rays with over 100 different energies
between 0.12 and 1.53 MeV, but the overall g ray emission rate is only about
1.5 g rays (of all energies) per decay of
152
Eu. The large majority of these
energies can be clearly resolved although only eleven energies are emitted
with suf®cient intensity (>3% each) to be useful for intensity calibrations. The
remaining nearly 90 energies are emitted with, on average less than 0.03%
probability each, so explaining the many very small peaks seen in Figure
3.14(a).
There are many other effects which could cause distortions in gamma ray
spectra, notably Compton backscatter. This is illustrated in Figure 3.15, with
details to come in Section 3.9.1.
Multi gamma ray emitters are of particular interest for energy and intensity
calibrations of high-resolution detectors when it is often possible to effect
calibrations over the entire energy range of interest using a single multi g ray
emitter, with europium-152 as an outstanding example. Other long-lived
multi g ray emitters frequently used for detector calibrations are europium-
154 and, for the lower energy range, barium-133 (Table 8.1). It is desirable to
use long half life radionuclides for this purpose, to justify the work involved
in the calibration. By contrast, NaI(Tl) detectors are rarely able to adequately
resolve more than a few of the peaks of multi g ray emitters (Figure 3.9(c))
except when used by specialists.
The lower energy section of spectra due to multi gamma ray sources is
invariably positioned on top of a large background due mainly to Compton
scatter. Pulses due to g rays which were Compton scattered out of the high
energy peaks build up below the lower energy part of the spectrum (Figures
3.10, 3.14(a) and 3.15). Background subtractions to obtain the net count in
full energy peaks are readily made for high-energy peaks but require
considerable care for the low-energy peaks since there are often many more
pulses in the background just below the peak than in the peak. This applies to
germanium detectors as much as to NaI(Tl) detectors.
For reasons yet to be described (Section 5.5.2), lithium drifted germanium
detectors cannot be prepared without dead layers which absorb all g rays
with energies below 30 keV (or sometimes below 40 keV) and cause signi®cant
attenuation of higher energy g rays up to about 100 keV. This is evident in
Figures 3.14(c) and 3.14(b) which respectively show the contrast between a
3.7 The role of mass energy 85